Experimental determination of the overall perturbation factor for the NACP chamber in electron beams fordmax<d⩽d80
In electron beam dosimetry with an ionization chamber, a factor that corrects for the cavity perturbation of the medium, Prepl, and one to account for the disturbance due to the chamber wall material differing from the medium, Pwall, are required. The overall perturbation correction factor, pq = PreplPwall, has been introduced because of the difficulty in separately measuring these two components. An advantage of parallel-plate ionization chambers is that pq has been shown to be close to unity at dmax. However, many dosimetry applications require knowledge of the overall perturbation factor at depths greater than dmax. We determined pq for the NACP chamber at depths beyond dmax by intercomparing percentage depth dose measurements made with it with those obtained with a PTW/diamond detector for which pq was taken as unity at all the measurement depths. Data were obtained at depths corresponding to approximately the 90 and 80 per cent of the dose maxima for 20, 16, 12 and 6 MeV incident electrons. The beam energy at depth, Ed, and the percentage depth-dose gradient varied from 1.4 to 14.3 MeV and 0 to 5.8% mm-1 respectively. Our results show that within the estimated uncertainty of 1.3%, pq,NACP is unity over the range of energies and dose gradients studied.
- 25 January 2001
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 46 (2) , N49-N55
Experimental determination of the overall perturbation factor for the NACP chamber in electron beams fordmax<d⩽d80
In electron beam dosimetry with an ionization chamber, a factor that corrects for the cavity perturbation of the medium, Prepl, and one to account for the disturbance due to the chamber wall material differing from the medium, Pwall, are required. The overall perturbation correction factor, pq = PreplPwall, has been introduced because of the difficulty in separately measuring these two components. An advantage of parallel-plate ionization chambers is that pq has been shown to be close to unity at dmax. However, many dosimetry applications require knowledge of the overall perturbation factor at depths greater than dmax. We determined pq for the NACP chamber at depths beyond dmax by intercomparing percentage depth dose measurements made with it with those obtained with a PTW/diamond detector for which pq was taken as unity at all the measurement depths. Data were obtained at depths corresponding to approximately the 90 and 80 per cent of the dose maxima for 20, 16, 12 and 6 MeV incident electrons. The beam energy at depth, Ed, and the percentage depth-dose gradient varied from 1.4 to 14.3 MeV and 0 to 5.8% mm-1 respectively. Our results show that within the estimated uncertainty of 1.3%, pq,NACP is unity over the range of energies and dose gradients studied.
- 12 February 1999
- Published by Wiley in Medical Physics
- Vol. 26 (2) , 208-213
Measured overall perturbation factors at depths greater than for ionization chambers in electron beams
In electron beam dosimetry the perturbation effect in the medium by the ionization chamber cavity is accounted for by introducing a replacement correction factor, Another perturbation correction factor, denoted as is due to the materials of the walls of the parallel‐plate chamber differing from the phantom material. Because of the difficulties in separating these two components, we measure the overall perturbation factor, A distinct advantage of parallel‐plate ionization chambers over cylindrical chambers is that has been shown to be close to unity at the standard calibration depth, However, for many dosimetry applications it is necessary to know the overall perturbation factor at depths greater than We measured the overall perturbation factor at depths greater than (approximating the 95%, 90% and 50% depth dose) for a Farmer‐type cylindrical ionization chamber and three parallel‐plate ionization chambers. We assumed that for the NACP chamber is unity at these measurement depths. The depth dependence for the other chambers was then measured relative to the NACP chamber. The mean energy at depth, and percentage depth dose gradient ranges studied were 1.9–18.5 MeV and 0 to 4.5%/mm, respectively. For the other two parallel‐plate chambers, we find to be unity at depths where the percent depth dose is greater than 90%, but it deviates from unity at deeper depths, where the dose gradients exceed about 2.5%/mm. For the cylindrical chamber, values at depths greater than were found to be in good agreement with those in TG 21, where the energy at depth, is used to evaluate
- 12 February 1999
- Published by Wiley in Medical Physics
- Vol. 26 (2) , 208-213
Measured overall perturbation factors at depths greater than for ionization chambers in electron beams
In electron beam dosimetry the perturbation effect in the medium by the ionization chamber cavity is accounted for by introducing a replacement correction factor, Another perturbation correction factor, denoted as is due to the materials of the walls of the parallel‐plate chamber differing from the phantom material. Because of the difficulties in separating these two components, we measure the overall perturbation factor, A distinct advantage of parallel‐plate ionization chambers over cylindrical chambers is that has been shown to be close to unity at the standard calibration depth, However, for many dosimetry applications it is necessary to know the overall perturbation factor at depths greater than We measured the overall perturbation factor at depths greater than (approximating the 95%, 90% and 50% depth dose) for a Farmer‐type cylindrical ionization chamber and three parallel‐plate ionization chambers. We assumed that for the NACP chamber is unity at these measurement depths. The depth dependence for the other chambers was then measured relative to the NACP chamber. The mean energy at depth, and percentage depth dose gradient ranges studied were 1.9–18.5 MeV and 0 to 4.5%/mm, respectively. For the other two parallel‐plate chambers, we find to be unity at depths where the percent depth dose is greater than 90%, but it deviates from unity at deeper depths, where the dose gradients exceed about 2.5%/mm. For the cylindrical chamber, values at depths greater than were found to be in good agreement with those in TG 21, where the energy at depth, is used to evaluate