Masashi Takada
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Masashi Takada has a total of 50 co-authors. Here are the top co-authors by number of shared publications:
Kei Aoyama
Industry Business Group Fuji Electric Co., Ltd. Hino Tokyo Japan
5 shared publicationsTakashi Nakamura
Industry Business Group Fuji Electric Co., Ltd. Hino Tokyo Japan
5 shared publicationsTomoya Nunomiya
Industry Business Group Fuji Electric Co., Ltd. Hino Tokyo Japan
5 shared publicationsAkihiko Masuda
National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba, Ibaraki, Japan
3 shared publicationsMasaru Nakamura
3 shared publicationsMasataka Narita
3 shared publications
Latest Publications
- 27 November 2025
- Published by Elsevier in Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
- Vol. 1084
Measurement of in-phantom thermal neutron spatial distributions utilizing aluminum-based indirect computed radiography for accelerator-based neutron capture therapy
- 27 November 2025
- Published by Elsevier in Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
- Vol. 1084
Measurement of in-phantom thermal neutron spatial distributions utilizing aluminum-based indirect computed radiography for accelerator-based neutron capture therapy
- 17 February 2025
- Published by Wiley in Medical Physics
- Vol. 52 (5) , 3298-3308
Novel approach to long‐term monitoring of accelerator‐based boron neutron capture therapy
Background: Boron neutron capture therapy (BNCT) was conducted in a hospital using an accelerator‐based neutron source. The neutron beam intensity at the patient position was evaluated offline using a gold‐based neutron activation method. During BNCT neutron beam irradiation on patients, the neutron intensity was controlled in real time by measuring the proton beam current irradiated on a lithium neutron target. The neutron intensity at NCCH decreased owing to the degradation of the lithium neutron target during neutron irradiation. The reduction in the neutron beam intensity could not be monitored via proton beam measurement due to the dependence of neutron production on the neutron target condition. Purpose: The duration of BNCT neutron irradiation should be controlled by monitoring the neutron beam intensity with a real‐time neutron detector for reliable neutron irradiation on patients. The measurement accuracy of the online neutron beam monitor was experimentally obtained by comparing the gold radioactivity measured at the patient position. Radiation‐induced damage was observed from the variation in the pulse height distributions of multichannel analyzer during long‐term neutron exposure. Methods: Neutron beams were measured during neutron beam irradiation at the BNCT facility of Edogawa hospital in Japan using a neutron beam monitor comprising a 0.07‐ LiF layer and 40‐ back‐illuminated thin Si pin diode. The proton beam was continuously irradiated until a cumulative total beam charge of approximately 3 kC was achieved. The online neutron beam monitor counting rates on the neutron target unit and gold saturation activities at the patient position were simultaneously measured through the entire duration of proton beam irradiation. Results: The experimental results demonstrated the long‐term operation of the online neutron beam monitor positioned on the neutron target unit during the entire duration of the neutron target lifespan without significant performance deterioration. A good synchronization was observed in a correlation distribution measured using the online neutron beam monitor and the gold neutron activation method. A conversion coefficient of 1.199 g with a standard deviation of 2.5% was evaluated. The neutron beam intensity irradiating on patients within an acceptable level of 5% as per the International Commission on Radiation Units and Measurements was evaluated from the online neutron counting rate at the 95% confidence level. The channel numbers of the triton peak and alpha particle edge decreased linearly owing to displacement damage and total ionizing dose effects induced mainly by thermal neutrons and photons. Conclusions: Neutron doses can be accurately administered by complementing proton beam current measurements with the online neutron beam monitor. The online neutron beam monitoring technique allows monitoring fluctuations in the neutron beam intensity and tracking the degradation of the lithium target through the neutron target lifespan. Using a calibrated online neutron beam monitor, a prescribed dose can be administered in a manner similar to that in x‐ray therapy, and the duration of neutron beam irradiation on the patient can be controlled in real time.
- 17 February 2025
- Published by Wiley in Medical Physics
- Vol. 52 (5) , 3298-3308
Novel approach to long‐term monitoring of accelerator‐based boron neutron capture therapy
Background: Boron neutron capture therapy (BNCT) was conducted in a hospital using an accelerator‐based neutron source. The neutron beam intensity at the patient position was evaluated offline using a gold‐based neutron activation method. During BNCT neutron beam irradiation on patients, the neutron intensity was controlled in real time by measuring the proton beam current irradiated on a lithium neutron target. The neutron intensity at NCCH decreased owing to the degradation of the lithium neutron target during neutron irradiation. The reduction in the neutron beam intensity could not be monitored via proton beam measurement due to the dependence of neutron production on the neutron target condition. Purpose: The duration of BNCT neutron irradiation should be controlled by monitoring the neutron beam intensity with a real‐time neutron detector for reliable neutron irradiation on patients. The measurement accuracy of the online neutron beam monitor was experimentally obtained by comparing the gold radioactivity measured at the patient position. Radiation‐induced damage was observed from the variation in the pulse height distributions of multichannel analyzer during long‐term neutron exposure. Methods: Neutron beams were measured during neutron beam irradiation at the BNCT facility of Edogawa hospital in Japan using a neutron beam monitor comprising a 0.07‐ LiF layer and 40‐ back‐illuminated thin Si pin diode. The proton beam was continuously irradiated until a cumulative total beam charge of approximately 3 kC was achieved. The online neutron beam monitor counting rates on the neutron target unit and gold saturation activities at the patient position were simultaneously measured through the entire duration of proton beam irradiation. Results: The experimental results demonstrated the long‐term operation of the online neutron beam monitor positioned on the neutron target unit during the entire duration of the neutron target lifespan without significant performance deterioration. A good synchronization was observed in a correlation distribution measured using the online neutron beam monitor and the gold neutron activation method. A conversion coefficient of 1.199 g with a standard deviation of 2.5% was evaluated. The neutron beam intensity irradiating on patients within an acceptable level of 5% as per the International Commission on Radiation Units and Measurements was evaluated from the online neutron counting rate at the 95% confidence level. The channel numbers of the triton peak and alpha particle edge decreased linearly owing to displacement damage and total ionizing dose effects induced mainly by thermal neutrons and photons. Conclusions: Neutron doses can be accurately administered by complementing proton beam current measurements with the online neutron beam monitor. The online neutron beam monitoring technique allows monitoring fluctuations in the neutron beam intensity and tracking the degradation of the lithium target through the neutron target lifespan. Using a calibrated online neutron beam monitor, a prescribed dose can be administered in a manner similar to that in x‐ray therapy, and the duration of neutron beam irradiation on the patient can be controlled in real time.
- 27 February 2023
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 70 (4) , 737-745
Response Function of Active Thermal Neutron Detector for Boron Neutron Capture Therapy Using Back-Illuminated Thin Si PIN Diode
A lithium fluoride-silicon-based neutron detector was developed for real-time measurement of neutron beams for boron neutron capture therapy (BNCT). Detection efficiencies for thermal neutrons were measured at the national standard neutron field in Japan and simulated using a Monte Carlo simulation code to evaluate the absolute BNCT neutron fluence rates. The simulation neutron detection efficiencies agreed excellently with the experimental results, within an experimental error of 3%. The simulation results reproduced well the peaks of tritons and alpha particles emitted from the 6 Li(n,t) 4 He reaction, peak of the protons produced by the 14 N(n,p) 14 C reaction in air, and distribution of alpha particles produced by the neutron capture reactions of dopant boron. The active BNCT neutron detector was characterized based on the experimental and simulation results.
- 27 February 2023
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 70 (4) , 737-745
Response Function of Active Thermal Neutron Detector for Boron Neutron Capture Therapy Using Back-Illuminated Thin Si PIN Diode
A lithium fluoride-silicon-based neutron detector was developed for real-time measurement of neutron beams for boron neutron capture therapy (BNCT). Detection efficiencies for thermal neutrons were measured at the national standard neutron field in Japan and simulated using a Monte Carlo simulation code to evaluate the absolute BNCT neutron fluence rates. The simulation neutron detection efficiencies agreed excellently with the experimental results, within an experimental error of 3%. The simulation results reproduced well the peaks of tritons and alpha particles emitted from the 6 Li(n,t) 4 He reaction, peak of the protons produced by the 14 N(n,p) 14 C reaction in air, and distribution of alpha particles produced by the neutron capture reactions of dopant boron. The active BNCT neutron detector was characterized based on the experimental and simulation results.
- 1 October 2021
- Published by Elsevier in Applied Radiation and Isotopes
- Vol. 176, 109856
Development of a real-time neutron beam detector for boron neutron capture therapy using a thin silicon sensor
- 1 October 2021
- Published by Elsevier in Applied Radiation and Isotopes
- Vol. 176, 109856
Development of a real-time neutron beam detector for boron neutron capture therapy using a thin silicon sensor
- 22 December 2017
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 180 (1-4) , 372-376
SIMULATED 8 MeV NEUTRON RESPONSE FUNCTIONS OF A THIN SILICON NEUTRON SENSOR
Neutron response functions of a thin silicon neutron sensor are simulated using PHITS2 and MCNP6 codes for an 8 MeV neutron beam at angles of incidence of 0°, 30° and 60°. The contributions of alpha particles created from the 28Si(n,α)25Mg reaction and the silicon nuclei scattered elastically by neutrons in the silicon sensor have not been well reproduced using the MCNP6 code. The 8 MeV neutron response functions simulated using the PHITS2 code with an accurate event generator mode are in good agreement with experimental results and include the contributions of the alpha particles and silicon nuclei.
- 22 December 2017
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 180 (1-4) , 372-376
SIMULATED 8 MeV NEUTRON RESPONSE FUNCTIONS OF A THIN SILICON NEUTRON SENSOR
Neutron response functions of a thin silicon neutron sensor are simulated using PHITS2 and MCNP6 codes for an 8 MeV neutron beam at angles of incidence of 0°, 30° and 60°. The contributions of alpha particles created from the 28Si(n,α)25Mg reaction and the silicon nuclei scattered elastically by neutrons in the silicon sensor have not been well reproduced using the MCNP6 code. The 8 MeV neutron response functions simulated using the PHITS2 code with an accurate event generator mode are in good agreement with experimental results and include the contributions of the alpha particles and silicon nuclei.
- 5 May 2016
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 170 (1-4) , 213-217
CHARACTERIZATION OF A THIN SILICON SENSOR FOR ACTIVE NEUTRON PERSONAL DOSEMETERS
A thin silicon sensor has been developed for active neutron personal dosemeters for use by aircrews and first responders. This thin silicon sensor is not affected by the funneling effect, which causes detection of cosmic protons and over-response to cosmic neutrons. There are several advantages to the thin silicon sensor: a decrease in sensitivity to gamma rays, an improvement of the energy detection limit for neutrons down to 0.8 MeV and an increase in the sensitivity to fast neutrons. Neutron response functions were experimentally obtained using 2.5 and 5 MeV monoenergy neutron beams and a 252Cf neutron source. Simulation results using the Monte Carlo N-Particle transport code agree quite well with the experimental ones when an energy deposition region shaped like a circular truncated cone is used in place of a cylindrical region.
- 5 May 2016
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 170 (1-4) , 213-217
CHARACTERIZATION OF A THIN SILICON SENSOR FOR ACTIVE NEUTRON PERSONAL DOSEMETERS
A thin silicon sensor has been developed for active neutron personal dosemeters for use by aircrews and first responders. This thin silicon sensor is not affected by the funneling effect, which causes detection of cosmic protons and over-response to cosmic neutrons. There are several advantages to the thin silicon sensor: a decrease in sensitivity to gamma rays, an improvement of the energy detection limit for neutrons down to 0.8 MeV and an increase in the sensitivity to fast neutrons. Neutron response functions were experimentally obtained using 2.5 and 5 MeV monoenergy neutron beams and a 252Cf neutron source. Simulation results using the Monte Carlo N-Particle transport code agree quite well with the experimental ones when an energy deposition region shaped like a circular truncated cone is used in place of a cylindrical region.
- 7 August 2011
- Published by Pleiades Publishing Ltd in Cosmic Research
- Vol. 49 (4) , 319-324
Evaluation of dose rate reduction in a spacecraft compartment due to additional water shield
The dose reduction rates brought about by the installation of additional water shielding in a spacecraft are calculated in the paper using the particles and heavy ion transport code system PHITS,...
- 7 August 2011
- Published by Pleiades Publishing Ltd in Cosmic Research
- Vol. 49 (4) , 319-324
Evaluation of dose rate reduction in a spacecraft compartment due to additional water shield
The dose reduction rates brought about by the installation of additional water shielding in a spacecraft are calculated in the paper using the particles and heavy ion transport code system PHITS,...
- 1 August 2004
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 110 (1-4) , 641-644
Microdosimetry of neutron field for boron neutron capture therapy at Kyoto university reactor
Microdosimetric single event spectrum in a human body simulated by an acrylic phantom has been measured for the clinical BNCT field at the Kyoto University Reactor (KUR). The recoil particles resulting from the initial reaction and subsequent interactions, namely protons, electrons, alpha particles and carbon nuclei are identified in the microdosimetric spectrum. The relative contributions to the neutron dose from proton, alpha particles and carbon are estimated to be about 0.9, 0.07 and 0.3, respectively, four depths between 5 and 41 mm. We estimate that the dose averaged lineal energy, yD decreased with depth from 64 to 46 keV µm−1. Relative biological effectiveness (RBE) of this neutron field using a response function for the microdosimetric spectrum was estimated to decrease from 3.6 to 2.9 with increasing depth.
- 1 August 2004
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 110 (1-4) , 641-644
Microdosimetry of neutron field for boron neutron capture therapy at Kyoto university reactor
Microdosimetric single event spectrum in a human body simulated by an acrylic phantom has been measured for the clinical BNCT field at the Kyoto University Reactor (KUR). The recoil particles resulting from the initial reaction and subsequent interactions, namely protons, electrons, alpha particles and carbon nuclei are identified in the microdosimetric spectrum. The relative contributions to the neutron dose from proton, alpha particles and carbon are estimated to be about 0.9, 0.07 and 0.3, respectively, four depths between 5 and 41 mm. We estimate that the dose averaged lineal energy, yD decreased with depth from 64 to 46 keV µm−1. Relative biological effectiveness (RBE) of this neutron field using a response function for the microdosimetric spectrum was estimated to decrease from 3.6 to 2.9 with increasing depth.