Paula Weidinger
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Paula Weidinger has a total of 23 co-authors. Here are the top co-authors by number of shared publications:
Gisa Foyer
Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany
2 shared publicationsRolf Kumme
Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany
2 shared publications- Alexander Dubowik
Physikalisch-Technische Bundesanstalt, Braunschweig, Germany
1 shared publication - Anna Trych-Wildner
Central Office of Measures (GUM)
1 shared publication C Schlegel
Physikalisch-Technische Bundesanstalt, Braunschweig, Germany
1 shared publication- Christian Mester
Federal Institute of Metrology METAS, Lindenweg 50, 3003 Bern-Wabern, Switzerland
1 shared publication
Latest Publications
- Open Access
Estimation of Uncertainty for the Torque Transducer in MNm Range—Classical Approach and Fuzzy Sets
The article provides an analysis of the metrological properties of a 5 MN·m torque transducer. The relative electrical signal (given in electrical units, mV/V) as a function of torque measured in both the clockwise and anticlockwise directions was monitored. To fit the data, the weighted method of least squares with both a straight-line and a cubic spline curve was used. The results of the analysis indicated that the straight-line fitting method produced smaller values of expanded uncertainty than the cubic spline fitting method. Additionally, the study confirmed the assumptions that the Monte Carlo method for propagating uncorrelated distributions was more accurate than the uncertainty propagation method, regardless of the type of curve fitting used. From the estimated uncertainty coverage corridor at selected measurement points, confidence intervals–expanded uncertainties were determined. Additionally, the fuzzy sets approach to the evaluation of uncertainty was presented, and the approximate value of the expanded uncertainty was calculated.
Open AccessEstimation of Uncertainty for the Torque Transducer in MNm Range—Classical Approach and Fuzzy Sets
The article provides an analysis of the metrological properties of a 5 MN·m torque transducer. The relative electrical signal (given in electrical units, mV/V) as a function of torque measured in both the clockwise and anticlockwise directions was monitored. To fit the data, the weighted method of least squares with both a straight-line and a cubic spline curve was used. The results of the analysis indicated that the straight-line fitting method produced smaller values of expanded uncertainty than the cubic spline fitting method. Additionally, the study confirmed the assumptions that the Monte Carlo method for propagating uncorrelated distributions was more accurate than the uncertainty propagation method, regardless of the type of curve fitting used. From the estimated uncertainty coverage corridor at selected measurement points, confidence intervals–expanded uncertainties were determined. Additionally, the fuzzy sets approach to the evaluation of uncertainty was presented, and the approximate value of the expanded uncertainty was calculated.
- Open Access
- 20 March 2023
- Published by Springer Nature in Forschung im Ingenieurwesen
- Vol. 87 (1) , 259-273
Traceable efficiency determination of a 2.75 MW nacelle on a test bench
In this paper, the efficiency of a 2.75 MW nacelle drivetrain on a test bench was determined traceably at various load points. For this purpose, so-called transfer standards for mechanical and electrical power measurement were additionally installed and integrated into the nacelle drivetrain. As torque measurement contributes greatly to the overall uncertainty, the static torque calibration was expanded to include additional influences present under rotation. An overall system efficiency of 89% was measured in the high torque and high speed range. The relative expanded measurement uncertainty for efficiency determination was between 0.30% and 0.72% over the entire operating range. Both the efficiency and the relative expanded measurement uncertainty were calculated for each operating point.
- 20 March 2023
- Published by Springer Nature in Forschung im Ingenieurwesen
- Vol. 87 (1) , 259-273
Open AccessTraceable efficiency determination of a 2.75 MW nacelle on a test bench
In this paper, the efficiency of a 2.75 MW nacelle drivetrain on a test bench was determined traceably at various load points. For this purpose, so-called transfer standards for mechanical and electrical power measurement were additionally installed and integrated into the nacelle drivetrain. As torque measurement contributes greatly to the overall uncertainty, the static torque calibration was expanded to include additional influences present under rotation. An overall system efficiency of 89% was measured in the high torque and high speed range. The relative expanded measurement uncertainty for efficiency determination was between 0.30% and 0.72% over the entire operating range. Both the efficiency and the relative expanded measurement uncertainty were calculated for each operating point.
- Open Access
- 1 January 2023
- Published by AMA Service GmbH
- p. 223-224
D4.2 - Zero Signal Determination for Torque Measurement Under Rotation in Test Benches
This paper discusses the problem of determining the zero point when measuring torque under rotation. Two approaches are presented and analysed: static zero signal determination, which should be used for efficiency calculation, and rotational zero signal determination, which should be used for test bench torque calibration under rotation.
- 1 January 2023
- Published by AMA Service GmbH
- p. 223-224
Open AccessD4.2 - Zero Signal Determination for Torque Measurement Under Rotation in Test Benches
This paper discusses the problem of determining the zero point when measuring torque under rotation. Two approaches are presented and analysed: static zero signal determination, which should be used for efficiency calculation, and rotational zero signal determination, which should be used for test bench torque calibration under rotation.
- Open Access
- 17 April 2019
- Published by Copernicus GmbH in Journal of Sensors and Sensor Systems
- Vol. 8 (1) , 149-159
Calibration of torque measurement under constant rotation in a wind turbine test bench
To determine the efficiency of multi-MW wind turbines, the torque measurement in wind turbine test benches has to be performed with an accuracy better than 0.2 %. To this end, the torque measurement has to be traced to national standards in the MN m range. This can be done by using a novel torque transfer standard in combination with a newly developed torque calibration procedure under constant rotation, which are both presented and discussed in this paper. The calibration procedure was performed on a 4 MW wind turbine test bench and the evaluation of the calibration result, including a measurement uncertainty, is explained in detail.
- 17 April 2019
- Published by Copernicus GmbH in Journal of Sensors and Sensor Systems
- Vol. 8 (1) , 149-159
Open AccessCalibration of torque measurement under constant rotation in a wind turbine test bench
To determine the efficiency of multi-MW wind turbines, the torque measurement in wind turbine test benches has to be performed with an accuracy better than 0.2 %. To this end, the torque measurement has to be traced to national standards in the MN m range. This can be done by using a novel torque transfer standard in combination with a newly developed torque calibration procedure under constant rotation, which are both presented and discussed in this paper. The calibration procedure was performed on a 4 MW wind turbine test bench and the evaluation of the calibration result, including a measurement uncertainty, is explained in detail.
- Open Access
- 1 August 2018
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 1065 (4) , 042057
Investigations towards extrapolation approaches for torque transducer characteristics
For torque transducers above 1.1 MNm, no traceable measurement has been possible to date. Therefore, an extrapolation method deploying measurement data from partial ranges is to be developed. Both partial and full range measurements up to 20 kN m are performed to gather data for implementing an extrapolation approach. The extrapolated data is evaluated, and the results are compared to the full range results.
- 1 August 2018
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 1065 (4) , 042057
Open AccessInvestigations towards extrapolation approaches for torque transducer characteristics
For torque transducers above 1.1 MNm, no traceable measurement has been possible to date. Therefore, an extrapolation method deploying measurement data from partial ranges is to be developed. Both partial and full range measurements up to 20 kN m are performed to gather data for implementing an extrapolation approach. The extrapolated data is evaluated, and the results are compared to the full range results.
- Open Access
- 20 April 2017
- Published by National Scientific Centre Institute of Metrology in Ukrainian Metrological Journal
- No. 1A,p. 57-60
Extending the Torque Calibration Range – Necessity and Outline of a Mathematical Approach
The current torque calibration range up to 1.1 MN·m is not any longer sufficient. The fast growing wind energy sector raised the demand for traceability of torque in the MN·m range over the last few years. To this end, a 5 MN·m torque transducer was acquired to extend the torque calibration range. Since the measuring device cannot be calibrated over its full measurement range, mathematical methods to forecast the behaviour of the torque transducer above 1.1 MN·m and their further development are presented.
- 20 April 2017
- Published by National Scientific Centre Institute of Metrology in Ukrainian Metrological Journal
- No. 1A,p. 57-60
Open AccessExtending the Torque Calibration Range – Necessity and Outline of a Mathematical Approach
The current torque calibration range up to 1.1 MN·m is not any longer sufficient. The fast growing wind energy sector raised the demand for traceability of torque in the MN·m range over the last few years. To this end, a 5 MN·m torque transducer was acquired to extend the torque calibration range. Since the measuring device cannot be calibrated over its full measurement range, mathematical methods to forecast the behaviour of the torque transducer above 1.1 MN·m and their further development are presented.