Francesca Ryan-Beswick
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Francesca Ryan-Beswick has a total of 10 co-authors. Here are the top co-authors by number of shared publications:
Caitlin Kyle
Department of Clinical Biochemistry, Manchester University NHS Foundation Trust, Manchester, UK
1 shared publicationCarrie Chadwick
The Neuroscience Laboratories, The Walton Centre NHS Foundation Trust, Liverpool, UK
1 shared publicationFaustina Owusu
Department of Clinical Biochemistry, Manchester University NHS Foundation Trust, Manchester, UK
1 shared publicationJonathan Scargill
Department of Clinical Biochemistry, Northern Care Alliance NHS Foundation Trust, Oldham, UK
1 shared publicationLaura Parke
Department of Clinical Biochemistry, Manchester University NHS Foundation Trust, Manchester, UK
1 shared publicationLesley Tetlow
Department of Clinical Biochemistry, Manchester University NHS Foundation Trust, Manchester, UK
1 shared publication
Latest Publications
- 15 September 2026
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Comparability of total bilirubin methods for neonatal samples across 6 NHS trusts within North West England
Background: The National Institute for Health and Care Excellence (NICE) guideline: management of neonatal jaundice [CG98] outlines total bilirubin (TBil) thresholds for the treatment of neonatal jaundice. Despite wide analytical variation in TBil, evidenced on external quality assessment schemes, the thresholds are based on a single Roche diazo assay. Adoption of these thresholds may result in inappropriate treatment decisions depending on the analytical methodology. This study assesses TBil method bias using pooled neonatal (<29 days) samples, prompted by incidents within Greater Manchester where differences in neonatal TBil measurements obtained due to transfer of care across sites caused clinical confusion. Methods: Seventeen serum and 9 plasma neonatal pools were analysed in duplicate on 6 platforms: Roche Cobas 701, Roche Pro, Abbott Alinity, Beckman Coulter AU5800, Siemens Atellica and Siemens XPT, utilising 5 diazo and 2 vanadate oxidation methodologies. Measurement biases compared to Roche Cobas were assessed. Results: Siemens Atellica and XPT vanadate oxidation methods exhibited significant positive bias: 17.4% [95% CI: 15.0–19.8%] and 17.0% [95% CI: 15.0–19.0%], respectively. All diazo methods were within desirable bias limits (+/−8%) as defined by the European Federation of Clinical Chemistry and Laboratory Medicine analytical performance specifications. Bias for Siemens vanadate oxidation and Beckman Coulter diazo methods exhibited matrix variation. Conclusions: Use of Siemens TBil vanadate oxidation method may lead to over treatment of neonatal jaundice when using NICE [CG98] thresholds. Laboratory professionals should evaluate the appropriateness of their TBil methodology in collaboration with their neonatal teams to minimise risks of overtreatment.
- 15 September 2026
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Comparability of total bilirubin methods for neonatal samples across 6 NHS trusts within North West England
Background: The National Institute for Health and Care Excellence (NICE) guideline: management of neonatal jaundice [CG98] outlines total bilirubin (TBil) thresholds for the treatment of neonatal jaundice. Despite wide analytical variation in TBil, evidenced on external quality assessment schemes, the thresholds are based on a single Roche diazo assay. Adoption of these thresholds may result in inappropriate treatment decisions depending on the analytical methodology. This study assesses TBil method bias using pooled neonatal (<29 days) samples, prompted by incidents within Greater Manchester where differences in neonatal TBil measurements obtained due to transfer of care across sites caused clinical confusion. Methods: Seventeen serum and 9 plasma neonatal pools were analysed in duplicate on 6 platforms: Roche Cobas 701, Roche Pro, Abbott Alinity, Beckman Coulter AU5800, Siemens Atellica and Siemens XPT, utilising 5 diazo and 2 vanadate oxidation methodologies. Measurement biases compared to Roche Cobas were assessed. Results: Siemens Atellica and XPT vanadate oxidation methods exhibited significant positive bias: 17.4% [95% CI: 15.0–19.8%] and 17.0% [95% CI: 15.0–19.0%], respectively. All diazo methods were within desirable bias limits (+/−8%) as defined by the European Federation of Clinical Chemistry and Laboratory Medicine analytical performance specifications. Bias for Siemens vanadate oxidation and Beckman Coulter diazo methods exhibited matrix variation. Conclusions: Use of Siemens TBil vanadate oxidation method may lead to over treatment of neonatal jaundice when using NICE [CG98] thresholds. Laboratory professionals should evaluate the appropriateness of their TBil methodology in collaboration with their neonatal teams to minimise risks of overtreatment.
- 5 June 2023
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 60 (5) , 353-355
The diagnostic utility of IgG index and oligoclonal bands for multiple sclerosis in a neurology hospital patient population
Background: Multiple sclerosis (MS) is a chronic demyelinating disease of the central nervous system. Diagnosis is based on the 2017 revised McDonald criteria. Unmatched oligoclonal bands (OCB) within the CSF (i.e. positive OCB) can substitute for dissemination in time by magnetic resonance imaging (MRI). Simonsen et al. (2020) claimed a raised (>0.7) immunoglobulin G (IgG) index could replace OCB status. This study aimed to establish the diagnostic utility of IgG index for MS in the population served by The Walton Centre NHS Foundation Trust (WCFT) a neurology and neurosurgery hospital, and to derive a population-based IgG index reference interval. Methods: OCB results from the laboratory information system (LIS) were collated from November 2018 to 2021. Final diagnosis and medication history was obtained from the electronic patient record. Exclusions were made based on age (0.7 is an appropriate cut-off to define a raised IgG index for the patient population.
- 5 June 2023
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 60 (5) , 353-355
The diagnostic utility of IgG index and oligoclonal bands for multiple sclerosis in a neurology hospital patient population
Background: Multiple sclerosis (MS) is a chronic demyelinating disease of the central nervous system. Diagnosis is based on the 2017 revised McDonald criteria. Unmatched oligoclonal bands (OCB) within the CSF (i.e. positive OCB) can substitute for dissemination in time by magnetic resonance imaging (MRI). Simonsen et al. (2020) claimed a raised (>0.7) immunoglobulin G (IgG) index could replace OCB status. This study aimed to establish the diagnostic utility of IgG index for MS in the population served by The Walton Centre NHS Foundation Trust (WCFT) a neurology and neurosurgery hospital, and to derive a population-based IgG index reference interval. Methods: OCB results from the laboratory information system (LIS) were collated from November 2018 to 2021. Final diagnosis and medication history was obtained from the electronic patient record. Exclusions were made based on age (0.7 is an appropriate cut-off to define a raised IgG index for the patient population.