Accuracy and precision of continuous, non‐invasive blood pressure devices: a systematic review and meta‐analyses
Open Access
- 15 June 2026
- journal article
- review article
- Published by Wiley in Anaesthesia
- Vol. 81 (11) , 1581-1592
Abstract
Summary: Introduction: Reliable continuous non‐invasive blood pressure monitoring technology should support clinicians to maintain blood pressure within appropriate limits peri‐operatively. Previous studies have shown conflicting results regarding measurement accuracy. Methods: This systematic review and six meta‐analyses evaluated the accuracy and precision of non‐invasive continuous arterial blood pressure measurement technologies (volume clamp technology, applanation tonometry and photoplethysmography) compared with invasive arterial measurement in adults. We analysed pooled bias (mean difference between non‐invasive and invasive measurement), standard deviation and limits of agreement. In each meta‐analysis, mean arterial pressure (MAP), systolic blood pressure (SBP) and diastolic blood pressure (DBP) were analysed separately. Acceptability was defined according to Association for the Advancement of Medical Instrumentation® (AAMI) criteria. Results: Sixty‐five studies were included: two meta‐analyses for volume clamp technology (ClearSight™, CNAP™); three meta‐analyses for applanation tonometry (T‐Line™, Vasotrac™ and N‐CAT™); and one meta‐analysis for photoplethysmography technology. For volume clamp technologies, none met AAMI criteria for bias and SD in relation to MAP, SBP and DBP. T‐Line, Vasotrac and N‐CAT met the AAMI criteria for bias for MAP (2.53 mmHg, 0.36 mmHg, ‐0.37 mmHg), SBP (‐0.07 mmHg, 1.20 mmHg, ‐3.68 mmHg) and DBP (3.09 mmHg, 1.09 mmHg, 0.84 mmHg), respectively. In terms of standard deviation, only N‐CAT (MAP 6.97 mmHg, SBP 7.14, mmHg, DBP 6.55 mmHg) met AAMI criteria for all blood pressure parameters. Photoplethysmography technology met AAMI criteria for bias (MAP ‐0.09 mmHg, SBP 0.49 mmHg, DBP ‐0.26 mmHg) and standard deviation (MAP 6.86 mmHg, DBP 6.38 mmHg). Discussion: Overall, the AAMI criteria for bias and SD were not fulfilled for most of the devices tested and wide limits of agreement are of concern. One applanation tonometry device (N‐CAT) shows promise. Clinicians should be aware of the lack of interchangeability with invasive blood pressure management.Keywords
This publication has 74 references indexed in Scilit:
- Association of intraoperative hypotension and cumulative norepinephrine dose with postoperative acute kidney injury in patients having noncardiac surgery: a retrospective cohort analysisBritish Journal of Anaesthesia, 2024
- PeriOperative Quality Initiative (POQI) international consensus statement on perioperative arterial pressure managementBritish Journal of Anaesthesia, 2024
- Accuracy and precision of continuous non-invasive arterial pressure monitoring in critical care: A systematic review and meta-analysisIntensive and Critical Care Nursing, 2021
- Updating guidance for reporting systematic reviews: development of the PRISMA 2020 statementJournal of Clinical Epidemiology, 2021
- Continuous noninvasive pulse wave analysis using finger cuff technologies for arterial blood pressure and cardiac output monitoring in perioperative and intensive care medicine: a systematic review and meta-analysisBritish Journal of Anaesthesia, 2020
- A framework for the meta‐analysis of Bland–Altman studies based on a limits of agreement approachStatistics in Medicine, 2017
- Continuous non-invasive arterial pressure measurement: Evaluation of CNAP™ device during vascular surgeryAnnales Françaises d'Anesthésie et de Réanimation, 2010
- Use of a radial artery compression device for noninvasive, near-continuous blood pressure monitoring in the EDThe American Journal of Emergency Medicine, 2004
- An evaluation of the N-CAT, a new arterial tonometerCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1995
- Assessment of the arterial tonometer (N-CAT) for the continuous blood pressure measurement in rapid atrial fibrillationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1993