Micro‐angiography for neuro‐vascular imaging. II. Cascade model analysis
- 24 October 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (11) , 3029-3039
Abstract
A micro‐angiographic detector was designed and its performance was previously tested to evaluate its feasibility as an improvement over current x‐ray detectors for neuro‐interventional imaging. The detector was shown to have a modulation transfer function value of about 2% at the Nyquist frequency of 10 cycles/mm and a zero frequency detective quantum efficiency [DQE(0)] value of about 55%. An assessment of the system was required to evaluate whether the current system was performing at its full potential and to determine if any of its components could be optimized to further improve the output. For the purpose, in this study, the parallel cascade theory was used to analyze the performance of the detector under neuro‐angiographic conditions by studying the output at the various stages in the imaging chain. A simple model for the spread of light in the CsI(Tl) entrance phosphor was developed and the resolution degradation due to ‐fluorescence absorption was calculated. The total gain of the system was found to result in (rms) detected at the charge coupled device per absorbed x‐ray photon. The gain and the spread of quanta in the imaging chain were used to calculate theoretically the DQE using the parallel cascade model. The results of the model‐based calculations matched fairly well with the experimental data previously obtained. This model was then used to optimize the phosphor thickness for the detector. The results showed that the area under the DQE curve had a maximum value at 150 μm of CsI(Tl), though when weighted by the squared signal in frequency space of a 100‐μm‐diam iodinated vessel, the integral DQE reached a maximum at 250 μm of CsI(Tl). Further, possible locations for gain increase in the imaging chain were determined, and the output of the improved system was simulated. Thus a theoretical analysis for the micro‐angiographic detector was performed to better assess its potential.Keywords
This publication has 15 references indexed in Scilit:
- Effects of characteristic x rays on the noise power spectra and detective quantum efficiency of photoconductive x‐ray detectorsMedical Physics, 2001
- Parallel cascades: New ways to describe noise transfer in medical imaging systemsMedical Physics, 2001
- Analysis of the detective quantum efficiency of a developmental detector for digital mammographyMedical Physics, 1999
- System considerations in CCD‐based x‐ray imaging for digital chest radiography and digital mammographyMedical Physics, 1997
- X‐ray imaging using amorphous selenium: Inherent spatial resolutionMedical Physics, 1995
- A spatial‐frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systemsMedical Physics, 1994
- Detective quantum efficiency of imaging systems with amplifying and scattering mechanismsJournal of the Optical Society of America A, 1987
- Energy and angular dependence of x-ray absorption and its effect on radiographic response in screen-film systemsPhysics in Medicine & Biology, 1983
- Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100Atomic Data and Nuclear Data Tables, 1970
- Random Noise Produced by X-Ray Fluorescent Screens*Journal of the Optical Society of America, 1968