Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system
Open Access
- 22 February 2025
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 26 (4) , e70057
Abstract
This article investigates the performance of a commercial BeO optically stimulated luminescent (OSL) dosimetry system (myOSLchip, RadPro GmbH International, Remscheid, Germany) through the application of the commissioning framework for luminescent dosimeters as described in the American Association of Physicists in Medicine Task Group 191 (AAPM TG191) report. Initial clinical experiences and dosimetric results are also presented. The following properties of the system were characterized: linearity correction factors ranged from ‐0.5% to +3% for dose levels spanning 0.1 to 20 Gy. Beam quality correction factors (relative to 6 MV) ranged from ‐4.5% (2.5FFF) to +4.5% (15MV) for photon beams and +1.9% (6 MeV) to +4.3% (20 MeV) for electron beams. An average (µ) signal loss per reading of ‐2.13% ± 0.20% was measured, however greater signal loss was observed in the first reading (µ = ‐2.6% ± 0.46%). An initial decline in individual element sensitivity relative to baseline was observed from 0–15 Gy cumulative dose (µ = ‐1.98% ± 0.55%), with negligible further deterioration from 15–32 Gy (µ = ‐2.38% ± 0.85%). Post‐irradiation, there was a transient OSL signal which faded with a half‐life of 1.8 min; this signal enhancement was +5% at 5 min post‐irradiation and +1% at 15 min relative to 24 h. Dosimeter response was not dependent on average dose rate in the range of 100–2500 MU/min. With respect to clinical testing, equal or superior performance compared with aluminum oxide OSLs (nanoDots) is shown for a range of clinical techniques and modalities including TSET, TBI, en‐face electrons, and pacemaker/out‐of‐field measurements. The feasibility of myOSLchip to serve as a primary clinical in vivo dosimetry system and direct replacement for Landauer's microStar system is demonstrated.Keywords
This publication has 22 references indexed in Scilit:
- A direct comparison of the optically stimulated luminescent properties of BeO and Al2O3 for clinical in-vivo dosimetryPhysical and Engineering Sciences in Medicine, 2022
- Out‐of‐field mean photon energy and dose from 6 MV and 6 MV FFF beams measured with TLD‐300 and TLD‐100 dosimetersMedical Physics, 2021
- Recumbent Total Skin Electron Beam TherapyAdvances in Radiation Oncology, 2021
- AAPM TG 191: Clinical use of luminescent dosimeters: TLDs and OSLDsMedical Physics, 2019
- Sensitivity and stability of optically stimulated luminescence dosimeters with filled deep electron/hole traps under pre-irradiation and bleaching conditionsPhysica Medica, 2017
- Optically stimulated luminescence of MgB 4 O 7 :Ce,Li for gamma and neutron dosimetryJournal of Luminescence, 2017
- Luminescence properties of BeO optically stimulated luminescence (OSL) detectorsRadiation Measurements, 2011
- Optically stimulated luminescence (OSL) dosimetry in medicinePhysics in Medicine & Biology, 2008
- Effect of high-dose irradiation on the optically stimulated luminescence of Al2O3:CRadiation Measurements, 2004
- OSL from BeO ceramics: new observations from an old materialRadiation Measurements, 1998