Electrogram Characteristics of Ablated and Non‐Ablated Myocardium in Humans: A Comparison of Miniaturized Embedded Electrodes and Conventional Ablation Electrodes
- 28 April 2016
- journal article
- clinical trial
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 27 (7) , 820-824
Abstract
Embedded Electrode Ablation: Introduction: Intracardiac electrogram voltage remains an important metric for radiofrequency lesion application. Embedded micro‐electrodes within ablation catheters are now approved for use in humans.Objective: This study examined electrogram characteristics of miniaturized electrode bipoles (iEGMmini) compared to distal ablation tip bipoles (iEGMstandard) in the setting of typical atrial flutter in humans.Methods: This was a single‐center prospective trial involving 21 patients. A total of 54 non‐ablated and 56 ablated sites were analyzed offline by separate blinded observers for iEGM voltage after a clinically successful radiofrequency (RF) ablation for typical atrial flutter. Ablation sites were defined as sites that were localized anatomically within the RF lesion set and did not have atrial capture at 10 mA and 2 milliseconds. Non‐ablated sites were defined as sites adjacent to or remote from the ablation lesion set with capture at 10 mA and 2 milliseconds.Results: iEGMmini had greater voltages at non‐ablated sites and lower voltages at ablated sites than iEGMstandard (non‐ablated sites iEGMstandard 0.64 mV vs. iEGMmini 1.0 mV; ablated sites iEGMstandard 0.15 mV, iEGMmini 0.08 mV, P < 0.001). The difference for each patient between ablated and non‐ablated sites was greater for iEGMmini compared to iEGMstandard (0.94 mV ± 0.57 iEGMmini vs. 0.52 mV ± 0.35 iEGMstandard, P < 0.001). Inter‐observer agreement was good among all non‐ablated sites and iEGMmini ablated sites, but was less robust for ablated sites using iEGMstandard.Conclusions: Imbedded mini‐electrode iEGM bipoles had higher voltages in non‐ablated tissue, lower voltages in ablated tissue, and significantly greater intra‐patient differences between ablated and non‐ablated sites. These data suggest a potential role for miniaturized electrode bipole use in RF lesion monitoring.Keywords
Funding Information
- Boston Scientific Corporation
This publication has 14 references indexed in Scilit:
- Maximal Electrogram Attenuation Recorded from Mini Electrodes Embedded on 4.5‐mm Irrigated and 8‐mm Nonirrigated Catheters Signifies Lesion MaturationJournal of Cardiovascular Electrophysiology, 2014
- 2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Patient Selection, Procedural Techniques, Patient Management and Follow-up, Definitions, Endpoints, and Research Trial DesignHeart Rhythm, 2012
- “Largest Amplitude Ablation” is the Optimal Approach for Typical Atrial Flutter Ablation: A Subanalysis from the AURUM 8 StudyJournal of Cardiovascular Electrophysiology, 2012
- Assessment of Disagreement: A New Information-Based ApproachAnnals of Epidemiology, 2010
- Functional Characterization of Atrial Electrograms in Sinus Rhythm Delineates Sites of Parasympathetic Innervation in Patients With Paroxysmal Atrial FibrillationJACC, 2007
- Electrogram Characteristics in Postinfarction Ventricular TachycardiaJACC, 2005
- Ablation Lesion Size Correlates with Pacing Threshold:Pacing and Clinical Electrophysiology, 2004
- Electrically Unexcitable Scar Mapping Based on Pacing Threshold for Identification of the Reentry Circuit IsthmusCirculation, 2002
- A critical discussion of intraclass correlation coefficientsStatistics in Medicine, 1994
- Electrophysiologic and anatomic basis for fractionated electrograms recorded from healed myocardial infarcts.Circulation, 1985