Oxygen and Glucose Distribution After Intracorneal Lens Implantation
- 1 December 2007
- journal article
- research article
- Published by Wiley in Optometry and Vision Science
- Vol. 84 (12) , 1074-1081
Abstract
Purpose.: Insertion of an implant in the cornea to achieve corneal multifocality has been suggested as a solution for presbyopia. However, unresolved issues related to nutrient transport need to be resolved. Our aim was to find the best lens position and influence lens transport properties in order to optimize nutrient supply to corneal cells. Method.: An axisymmetric corneal model was built to simulate the nutrient transport in the cornea. Oxygen and glucose concentrations were calculated for normal cornea and intracorneal lens wearing conditions. The simulation considers the different tissue layers (epithelium, stroma, and endothelium) as well as layer and solute concentration dependent consumption. Results.: The minimum oxygen tension in the cornea was found to be higher when the lens was placed at 3/4 of the corneal thickness. Moreover, in this position, the influence of the inlay diffusivity was smaller than at more anterior or posterior placements. The diffusivity of the inlay affects the way nutrients will be transported through the cornea. The threshold where glucose may diffuse through or around the implant was found to be 1/100th of the stromal diffusivity. Conclusions.: Computational methods are especially attractive to study nutrient transport in the cornea due to the difficulties associated with in vivo or in vitro measurements. The exact parameters that dictate the corneal metabolism are not known. However, the combined analysis of oxygen and glucose distribution is valuable in order to predict the complex physiological changes that arise under intracorneal lens implantation.Keywords
Funding Information
- National Science Foundation
This publication has 22 references indexed in Scilit:
- Two-Year Preclinical Testing of Perfluoropolyether Polymer as a Corneal InlayInvestigative Ophthalmology & Visual Science, 2006
- Corneal oxygenation during contact lens wear: comparison of diffusion and EOP‐based flux modelsClinical and Experimental Optometry, 2005
- Re-Evaluation of the Oxygen Diffusion Model for Predicting Minimum Contact Lens Dk/t Values Needed to Avoid Corneal AnoxiaOptometry and Vision Science, 1999
- A thin glycoprotein coating of a synthetic lenticule does not cause nutritional deficiency of the anterior corneaCurrent Eye Research, 1999
- Modeling Glucose Distribution in the CorneaCurrent Eye Research, 1990
- Complications of Hydrogel Intracorneal Lenses in MonkeysArchives of Ophthalmology (1950), 1987
- Oxygen tension distributions in the cornea: A re-examinationExperimental Eye Research, 1974
- Sugar transport across the corneal endotheliumExperimental Eye Research, 1969
- The Effect of Intralamellar Water-Impermeable Membranes on Corneal HydrationArchives of Ophthalmology (1950), 1966
- Effect of Intralamellar Plastic Membranes on Corneal PhysiologyAmerican Journal of Ophthalmology, 1961