A Nested Eight-Channel Transmit Array With Open-Face Concept for Human Brain Imaging at 7 Tesla
Authors:
Open Access
- 27 July 2021
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Physics
- Vol. 9
Abstract
Purpose: Parallel transmit technology for MRI at 7 tesla will significantly benefit from high performance transmit arrays that offer high transmit efficiency and low mutual coupling between the individual array elements. A novel dual-mode transmit array with nested array elements has been developed to support imaging the human brain in both the single-channel (sTx) and parallel-transmit (pTx) excitation modes of a 7 tesla MRI scanner. In this work, the design, implementation, validation, specific absorption rate (SAR) management, and performance of the head coil is presented.Methods: The transmit array consisted of a nested arrangement to improve decoupling between the second-neighboring elements. Two large cut-outs were introduced in the RF shield for an open-face design to reduce claustrophobia and to allow patient monitoring. A hardware interface allows the coil to be used in both the sTx and pTx modes. SAR monitoring is done with virtual observation points (VOP) derived from human body models. The transmit efficiency and coverage is compared with the commercial single-channel and parallel-transmit head coils.Results: Decoupling inductors between the second-neighboring coil elements reduced the coupling to less than −20 dB. Local SAR estimates from the electromagnetic (EM) simulations were always less than the EM-based VOPs, which in turn were always less than scanner predictions and measurements for static and dynamic pTx waveforms. In sTx mode, we demonstrate improved coverage of the brain compared to the commercial sTx coil. The transmit efficiency is within 10% of the commercial pTx coil despite the two large cut-outs in the RF shield. In pTx mode, improved signal homogeneity was shown when the Universal Pulse was used for acquisition in vivo.Conclusion: A novel head coil which includes a nested eight-channel transmit array has been presented. The large cut-outs improve patient monitoring and reduce claustrophobia. For pTx mode, the EM simulation and VOP-based SAR management provided greater flexibility to apply pTx methods without the limitations of SAR constraints. For scanning in vivo, the coil was shown to provide an improved coverage in sTx mode compared to a standard commercial head coil.Keywords
This publication has 38 references indexed in Scilit:
- A Specialized Multi-Transmit Head Coil for High Resolution fMRI of the Human Visual Cortex at 7TPLOS ONE, 2016
- Comparison of Routine Brain Imaging at 3 T and 7 TInvestigative Radiology, 2016
- A 16‐channel dual‐row transmit array in combination with a 31‐element receive array for human brain imaging at 9.4 TMagnetic Resonance in Medicine, 2013
- Transceiver-Phased Arrays for Human Brain Studies at 7 TApplied Magnetic Resonance, 2011
- A conformal transceive array for 7 T neuroimagingMagnetic Resonance in Medicine, 2011
- Design and evaluation of an RF front‐end for 9.4 T human MRIMagnetic Resonance in Medicine, 2011
- A radiofrequency coil to facilitate B shimming and parallel imaging acceleration in three dimensions at 7 TNMR in Biomedicine, 2010
- B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coilMagnetic Resonance in Medicine, 2005
- Transmit and receive transmission line arrays for 7 Tesla parallel imagingMagnetic Resonance in Medicine, 2005
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004