A novel nanohydroxyapatite/polyamide 66 strut for reducing subsidence after one-level anterior cervical corpectomy and fusion: a finite-element study
Open Access
- 19 July 2024
- journal article
- research article
- Published by Springer Nature in European Journal of Medical Research
- Vol. 29 (1) , 1-8
Abstract
Background: The aim of this study is to introduce a novel nanohydroxyapatite/polyamide 66(n-HA/PA66)n strut to improve biomechanical performance and reduce subsidence. Methods: One validated intact and 2 ACCF-simulated C3–C7 cervical spine models were developed (old strut: Group A, new strut: Group B). In the ACCF models, C5 underwent corpectomy and was fixed by an anterior cervical plate. Screw angles were categorized as 1 (0 ) and 2 (45 ) and divided into 4 groups, A1, A2, B1 and B2, for each model. An axial force of 74 N and a moment couple of 1.0 Nm were imposed on the C3 vertebra. The range of motion (ROM) of each segment and the stress distribution on the screw–vertebra interface, strut, and strut–endplate interface were recorded and analysed. Results: There was no significant difference in ROM between Group A and Group B during bending, extension and rotation under 74 N axial pressure. The stress concentration on the strut body in Group A was higher than that in Group B. The peak stress values at the screw–vertebral interface in Groups A1 and A2 were higher than those in Groups B1 and B2, except for during extension and lateral bending. Under axial pressure, the peak stress values at the strut body–endplate interface during bending, extension and rotation were lower in the A1 and A2 groups than in the B1 and B2 groups. The Group B model showed much higher graft stress than the Group A model. Conclusions: Based on finite-element analysis, compared with the old strut, the novel strut showed better biomechanical performance at the screw–vertebra interface.Keywords
Funding Information
- Chongqing Natural Science Foundation Project (CSTB2022NSCQ-MSX0070)
- the medical program of Chongqing health and science,technology commission (2021MSXM285)
This publication has 16 references indexed in Scilit:
- A novel anatomic titanium mesh cage for reducing the subsidence rate after anterior cervical corpectomy: a finite element studyScientific Reports, 2021
- A biomechanical analysis of four anterior cervical techniques to treating multilevel cervical spondylotic myelopathy: a finite element studyBMC Musculoskeletal Disorders, 2021
- Effects of New Cage Profiles on the Improvement in Biomechanical Performance of Multilevel Anterior Cervical Corpectomy and Fusion: A Finite Element AnalysisWorld Neurosurgery, 2019
- Nanohydroxyapatite/polyamide 66 strut subsidence after one-level corpectomy: underlying mechanism and effect on cervical neurological functionScientific Reports, 2018
- Comparison of Outcomes for Anterior Cervical Discectomy and Fusion With and Without Anterior Plate FixationSpine, 2018
- Corpectomy versus discectomy for the treatment of multilevel cervical spine pathology: a finite element model analysisThe Spine Journal, 2012
- Biomechanical effects of anterior, posterior, and combined anterior-posterior instrumentation techniques on the stability of a multilevel cervical corpectomy construct: a finite element model analysisThe Spine Journal, 2011
- Validation of a C2–C7 cervical spine finite element model using specimen-specific flexibility dataMedical Engineering & Physics, 2010
- Screw angulation affects bone-screw stresses and bone graft load sharing in anterior cervical corpectomy fusion with a rigid screw-plate construct: a finite element model studyThe Spine Journal, 2009
- Mechanical Properties of the Human Cervical Spine as Shown by Three-Dimensional Load–Displacement CurvesSpine, 2001