Feasibility Studies of a New Event Selection Method to Improve Spatial Resolution of Compton Imaging for Medical Applications
- 10 May 2017
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 1 (4) , 358-367
Abstract
A Compton imaging method for medical applications that includes new energy determination and data filtering techniques has been tested using several point sources with known emission lines. Using a prototype Compton camera, a distance-of-closest approach technique has been employed to determine the initial energy of the incoming ys and to ensure the reconstructed source position is within an acceptable distance from the known y source location. Further analysis is done by implementing a Compton line filtering technique, keeping only those interactions whose deposited energy in the first interaction matches the theoretical energy deposition predicted by the Compton equation. Using this new event filtering method, we see improvements in the full width at half maximum in the lateral profiles of y point sources of up to 70% over standard Compton imaging methods, as well as achievable spatial resolutions in the reconstructed images of better than 2 mm. In addition, this new Compton imaging method was able to reconstruct an extended source of y rays emitted during irradiation of a water tank with a clinical proton radiotherapy beam.Keywords
Funding Information
- National Cancer Institute of the National Institutes of Health (R01CA187416)
This publication has 27 references indexed in Scilit:
- Proton therapy monitoring by Compton imaging: influence of the large energy spectrum of the prompt-γradiationPhysics in Medicine & Biology, 2016
- Imaging of prompt gamma rays emitted during delivery of clinical proton beams with a Compton camera: feasibility studies for range verificationPhysics in Medicine & Biology, 2015
- Test of Compton camera components for prompt gamma imaging at the ELBE bremsstrahlung beamJournal of Instrumentation, 2014
- Hodoscope coincidence imaging for hadron therapy using a compton camera2011
- A Compton imaging algorithm for on-line monitoring in hadron therapy2011
- A Compton imager for in-vivo dosimetry of proton beams—A design studyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Design Guidelines for a Double Scattering Compton Camera for Prompt-$\gamma$ Imaging During Ion Beam Therapy: A Monte Carlo Simulation StudyIEEE Transactions on Nuclear Science, 2010
- A Tracking Compton-Scattering Imaging System for Hadron Therapy MonitoringIEEE Transactions on Nuclear Science, 2010
- Gamma-ray energy-imaging integrated spectral deconvolutionNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Three-Compton telescope: theory, simulations, and performanceIEEE Transactions on Nuclear Science, 2002