Constant linear velocity spiral scanning for near video rate 4D OCT ophthalmic and surgical imaging with isotropic transverse sampling
Authors:
- 28 September 2018
- journal article
- research article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 9 (10) , 5052-5070
Abstract
Ultrahigh speed optical coherence tomography (OCT) systems with >100 kHz A-scan rates can generate volumes rapidly with minimal motion artifacts and are well suited for 4D imaging (volumes through time) applications such as intra-operative imaging. In such systems, high OCT data acquisition efficiency (defined as the fraction of usable A-scans generated during the total acquisition time) is desired to maximize the volumetric frame rate and sampling pitch. However, current methods for beam scanning using non-resonant and resonant mirror scanners can result in severe scan distortion and transverse oversampling as well as require acquisition dead times, which limit the acquisition efficiency and performance of ultrahigh speed 4D OCT. We introduce constant linear velocity spiral scanning (CLV-SC) as a novel beam scanning method to maximize the data acquisition efficiency of ultrahigh speed 4D OCT systems. We demonstrate that CLV-SC does not require acquisition dead times and achieves more uniform transverse sampling compared to raster scanning. To assess its clinical utility, we implement CLV-SC with a 400 kHz OCT system and image the anterior eye and retina of healthy adults at up to 10 volumes per second with isotropic transverse sampling, allowing B-scans with equal sampling pitch to be extracted from arbitrary locations within a single volume. The feasibility of CLV-SC for intra-operative imaging is also demonstrated using a 800 kHz OCT system to image simulated retinal surgery at 15 volumes per second with isotropic transverse sampling, resulting in high quality volume renders that enable clear visualization of surgical instruments and manipulation of tissue.Keywords
Funding Information
- Ford Foundation Predoctoral Fellowship
- National Eye Institute (R01-EY023039, R01-EY24312)
- National Science Foundation Graduate Fellowship
- NIH T32 Training Grant
- U.S. Department of War (CDMRP Contract W81XWH-16-1-0498)
This publication has 29 references indexed in Scilit:
- Review of intraoperative optical coherence tomography: technology and applications [Invited]Biomedical Optics Express, 2017
- Live volumetric (4D) visualization and guidance of in vivo human ophthalmic surgery with intraoperative optical coherence tomographyScientific Reports, 2016
- Optical Coherence Tomography for Retinal Surgery: Perioperative Analysis to Real-Time Four-Dimensional Image-Guided SurgeryInvestigative Ophthalmology & Visual Science, 2016
- High definition live 3D-OCT in vivo: design and evaluation of a 4D OCT engine with 1 GVoxel/sBiomedical Optics Express, 2014
- Handheld ultrahigh speed swept source optical coherence tomography instrument using a MEMS scanning mirrorBiomedical Optics Express, 2013
- Spectral domain optical coherence tomography of multi-MHz A-scan rates at 1310 nm range and real-time 4D-display up to 41 volumes/secondBiomedical Optics Express, 2012
- Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasersBiomedical Optics Express, 2012
- High speed spectral domain optical coherence tomography for retinal imaging at 500,000 A‑lines per secondBiomedical Optics Express, 2011
- Real-time intraoperative 4D full-range FD-OCT based on the dual graphics processing units architecture for microsurgery guidanceBiomedical Optics Express, 2011
- Optical Coherence TomographyScience, 1991