Assessing potential for aortoiliac vascular injury from venoarterial extracorporeal membrane oxygenation cannulae: An in vitro particle image velocimetry study
- 31 August 2020
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 45 (2) , E14-E25
Abstract
Background Limb ischemia is a major complication associated with peripheral veno‐arterial extracorporeal membrane oxygenation (VA ECMO). The high velocity jet from arterial cannulae can cause “sandblasting” injuries to the arterial endothelium, with the potential risk of distal embolization and end organ damage. Aim The aim of this study was to identify, for a range of clinically relevant VA ECMO cannulae and flow rates, any regions of peak flow velocity on the aortic wall which may predispose to vascular injury, and any regions of low velocity flow which may predispose to thrombus formation. Methods A silicone model of the aortic and iliac vessels was sourced and the right external iliac artery was cannulated. Cannulae ranged from 15 – 21 Fr in size. Simulated steady state ECMO flow rates were instituted using a magnetically levitated pump (CentriMag pump). Adaptive particle image velocimetry was performed for each cannula at 3, 3.5, 4 and 4.5 L/min. Results For all cannulae, in both horizontal and vertical side hole orientations, the peak velocity on the aortic wall ranged from 0.3 m/s ‐ 0.45 m/s, and the regions of lowest velocity flow were 0.05 m/s. The magnitude of peak velocity flow on the aortic wall was not different between a single pair versus multiple pairs of side holes. Maximum velocity flow on the aortic wall occurred earlier at a lower pump flow rate in the vertical orientation of distal side holes compared to a horizontal position. The presence of multiple paired side holes was associated with fewer low velocity flow regions, and some retrograde flow, in the distal abdominal aorta compared to cannulae with a single pair of side holes. Conclusion From this in vitro visualization study, the selection of a cannula design with multiple versus single pairs of side holes did not change the magnitude of peak velocity flow delivered to the vessel wall. Cannulae with multiple side holes were associated with fewer regions of low velocity flow in the distal abdominal aorta. Further in vivo studies, and ideally clinical data would be required to assess any correlation of peak velocity flows with incidence of vascular injury, and any low velocity flow regions with incidence of thrombosis.Keywords
This publication has 22 references indexed in Scilit:
- Technical-Induced Hemolysis in Patients with Respiratory Failure Supported with Veno-Venous ECMO – Prevalence and Risk FactorsPLOS ONE, 2015
- Hydrodynamic evaluation of a new dispersive aortic cannula (Stealthflow)Journal of Artificial Organs, 2015
- Hydrodynamic evaluation of aortic cardiopulmonary bypass cannulae using particle image velocimetryPerfusion, 2015
- Complications of Extracorporeal Membrane Oxygenation for Treatment of Cardiogenic Shock and Cardiac Arrest: A Meta-Analysis of 1,866 Adult PatientsThe Annals of Thoracic Surgery, 2014
- Arterial Vascular Complications in Peripheral Extracorporeal Membrane Oxygenation Support: A Review of Techniques and OutcomesFuture Cardiology, 2013
- ECMO retrieval in NSW and beyondCurrent Anaesthesia & Critical Care, 2010
- The diameter of the common femoral artery in healthy human: Influence of sex, age, and body sizeJournal of Vascular Surgery, 1999
- The Björk-Shiley aortic prosthesis: flow characteristics, thrombus formation and tissue overgrowth.Circulation, 1978
- Measured Turbulence and Its Effect on Thrombus FormationCirculation Research, 1974
- STUDIES IN EXPERIMENTAL EXTRACORPOREAL THROMBOSISJournal of Experimental Medicine, 1927