Abstract
An ozone (O3) analyzer for use in eddy‐correlation measurements is described which has an unfiltered bandwidth of at least 20 Hz and sensitivity of at least 0.1 ppbv at 12‐Hz bandwidth. Based on chemiluminescence with nitric oxide, it is designed primarily for use on meteorological aircraft, but it also works well for tower measurements. It is significant that the detecting mechanism responds linearly to time‐dependent inputs, which is a definite advantage for applications involving measurements of eddy fluxes and spectra of O3. The unit is highly specific, the only known interference being a slight quenching of the chemiluminescence by water vapor. The unit does not respond to cloud liquid water or to precipitation, and excellent performance has been obtained in the presence of these two conditions. The entire package weighs 110 kg and requires 360 W at 110 V ac (60 or 400 Hz), and 1200 W at 28 V dc.

This publication has 15 references indexed in Scilit: