Effect of Pulse Polarity on Thresholds and on Non-monotonic Loudness Growth in Cochlear Implant Users
- 30 January 2017
- journal article
- research article
- Published by Springer Nature in Journal of the Association for Research in Otolaryngology
- Vol. 18 (3) , 513-527
Abstract
Most cochlear implants (CIs) activate their electrodes non-simultaneously in order to eliminate electrical field interactions. However, the membrane of auditory nerve fibers needs time to return to its resting state, causing the probability of firing to a pulse to be affected by previous pulses. Here, we provide new evidence on the effect of pulse polarity and current level on these interactions. In experiment 1, detection thresholds and most comfortable levels (MCLs) were measured in CI users for 100-Hz pulse trains consisting of two consecutive biphasic pulses of the same or of opposite polarity. All combinations of polarities were studied: anodic-cathodic-anodic-cathodic (ACAC), CACA, ACCA, and CAAC. Thresholds were lower when the adjacent phases of the two pulses had the same polarity (ACCA and CAAC) than when they were different (ACAC and CACA). Some subjects showed a lower threshold for ACCA than for CAAC while others showed the opposite trend demonstrating that polarity sensitivity at threshold is genuine and subject- or electrode-dependent. In contrast, anodic (CAAC) pulses always showed a lower MCL than cathodic (ACCA) pulses, confirming previous reports. In experiments 2 and 3, the subjects compared the loudness of several pulse trains differing in current level separately for ACCA and CAAC. For 40 % of the electrodes tested, loudness grew non-monotonically as a function of current level for ACCA but never for CAAC. This finding may relate to a conduction block of the action potentials along the fibers induced by a strong hyperpolarization of their central processes. Further analysis showed that the electrodes showing a lower threshold for ACCA than for CAAC were more likely to yield a non-monotonic loudness growth. It is proposed that polarity sensitivity at threshold reflects the local neural health and that anodic asymmetric pulses should preferably be used to convey sound information while avoiding abnormal loudness percepts.Keywords
Funding Information
- Agence Nationale de la Recherche (ANR-11-PDOC-0022)
- Medical Research Council (MC-A060-5PQ70)
- Centre National de la Recherche Scientifique (PICS)
This publication has 29 references indexed in Scilit:
- Effects of Pulse Shape and Polarity on Sensitivity to Cochlear Implant Stimulation: A Chronic Study in Guinea Pigs2016
- Re-examining the upper limit of temporal pitchThe Journal of the Acoustical Society of America, 2014
- Polarity effects on place pitch and loudness for three cochlear-implant designs and at different cochlear sitesThe Journal of the Acoustical Society of America, 2013
- Probing the Electrode-Neuron Interface With Focused Cochlear Implant StimulationTrends in Amplification, 2010
- Asymmetric Pulses in Cochlear Implants: Effects of Pulse Shape, Polarity, and RateJournal of the Association for Research in Otolaryngology, 2006
- Histopathology of Cochlear Implants in HumansAnnals of Otology, Rhinology & Laryngology, 2001
- The basic mechanism for the electrical stimulation of the nervous systemNeuroscience, 1999
- Electrically evoked compound action potentials of guinea pig and cat: responses to monopolar, monophasic stimulationHearing Research, 1998
- Effects of phase duration on detection of electrical stimulation of the human cochleaHearing Research, 1993
- Better speech recognition with cochlear implantsNature, 1991