Perfusion Changes at the Forehead Measured by Photoplethysmography during a Head-Down Tilt Protocol
Open Access
- 27 May 2019
- journal article
- research article
- Published by MDPI AG in Biosensors
- Vol. 9 (2) , 71
Abstract
Photoplethysmography (PPG) signals from the forehead can be used in pulse oximetry as they are less affected by vasoconstriction compared to fingers. However, the increase in venous blood caused by the positioning of the patient can deteriorate the signals and cause erroneous estimations of the arterial oxygen saturation. To date, there is no method to measure this venous presence under the PPG sensor. This study investigates the feasibility of using PPG signals from the forehead in an effort to estimate relative changes in haemoglobin concentrations that could reveal these posture-induced changes. Two identical reflectance PPG sensors were placed on two different positions on the forehead (above the eyebrow and on top of a large vein) in 16 healthy volunteers during a head-down tilt protocol. Relative changes in oxygenated ( HbO ), reduced ( HHb) and total ( tHb) haemoglobin were estimated from the PPG signals and the trends were compared with reference Near Infrared Spectroscopy (NIRS) measurements. Also, the signals from the two PPG sensors were analysed in order to reveal any difference due to the positioning of the sensor. HbO , HHb and tHb estimated from the forehead PPGs trended well with the same parameters from the reference NIRS. However, placing the sensor over a large vasculature reduces trending against NIRS, introduces biases as well as increases the variability of the changes in HHb. Forehead PPG signals can be used to measure perfusion changes to reveal venous pooling induced by the positioning of the subject. Placing the sensor above the eyebrow and away from large vasculature avoids biases and large variability in the measurements.
Keywords
Funding Information
- Barts and The London NHS Trust (Barts Charity) (832/1716)
This publication has 25 references indexed in Scilit:
- Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitationsRespiratory Medicine, 2013
- A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac OutputAnesthesia & Analgesia, 2010
- Forehead Reflectance Oximetry: A Clinical Comparison with Conventional Digit Sensors during Laparotomic and Laparoscopic Abdominal SurgeryJournal of Clinical Monitoring and Computing, 2007
- Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusionHeart & Lung, 2007
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Forehead Pulse OximetryAnesthesiology, 2006
- The Effect of Venous Pulsation on the Forehead Pulse Oximeter Wave Form as a Possible Source of Error in Spo2 CalculationAnesthesia & Analgesia, 2005
- The Influence of Larger Subcutaneous Blood Vessels on Pulse OximetryJournal of Clinical Monitoring and Computing, 2004
- In Vivo Near-Infrared SpectroscopyAnnual Review of Biomedical Engineering, 2000
- The detection of peripheral venous pulsation using the pulse oximeter as a plethysmographJournal of Clinical Monitoring and Computing, 1993