Abstract
G surface, and injury to the surgical team remain concerning. Questions/purposes We determined the rate of cable failure, union, and complications associated with a novel, nonmetallic cerclage cable in periprosthetic fracture and osteotomy fixation during THA. Methods We retrospectively reviewed 29 patients who had primary and revision THAs using nonmetallic cables. Indications for use included fixation of an extended trochanteric osteotomy, intraoperative fracture of the proximal femur, strut allograft fixation, and a Vancouver B1 periprosthetic fracture of the femur. All patients were evaluated clinically and radiographically immediately postoperatively, at 3 weeks, 6 weeks, 3 months, and then annually thereafter. The minimum followup was 13 months (mean, 21 months; range, 13-30 months). Results Two of the 29 patients (7%) developed a nonunion; all remaining osteotomies, fractures and allografts had healed at the time of most recent evaluation. Four patients (14%) dislocated postoperatively; two were treated successfully with closed reduction, while the other two required reoperation. We identified no evidence of breakage or other complications directly attributable to the cables. Conclusions The nonmetallic periprosthetic cables used in this series provided adequate fixation to allow for both osteotomy and fracture healing. We did not observe any complications directly related to the cables. Level of Evidence Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence....Keywords
This publication has 30 references indexed in Scilit:
- Cemented Total Hip Replacement Cable Debris and Acetabular Construct DurabilityJournal of Bone and Joint Surgery, 2009
- Use of the Extended Trochanteric Osteotomy in Treating Prosthetic Hip InfectionThe Journal of Arthroplasty, 2009
- Extended Trochanteric Osteotomy for the Treatment of Vancouver B2/B3 Periprosthetic Fractures of the FemurThe Journal of Arthroplasty, 2008
- Fixation Systems of Greater Trochanteric Osteotomies: Biomechanical and Clinical OutcomesJournal of the American Academy of Orthopaedic Surgeons, 2007
- A Clinical, Radiographic, and Cost Comparison of Cerclage Techniques: Wires vs CablesThe Journal of Arthroplasty, 2006
- Current Status of Trochanteric Reattachment in Complex Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2005
- The use of uncemented extensively porous-coated femoral components in the management of Vancouver B2 and B3 periprosthetic femoral fracturesThe Journal of Bone and Joint Surgery. British volume, 2005
- Periprosthetic Fractures Evaluation and TreatmentClinical Orthopaedics and Related Research, 2004
- The reliability of validity of the Vancouver classification of femoral fractures after hip replacementThe Journal of Arthroplasty, 2000
- Complications of a cable grip systemThe Journal of Arthroplasty, 1996