Unsuccessful long-term outcome after treatment of a vein graft false aneurysm with a polytetrafluoethylene-coated Jostent
- 12 May 2000
- journal article
- case report
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 50 (1) , 105-108
Abstract
We describe a case in which a stented ulcerated plaque in an old vein graft ruptured to a huge false aneurysm. By the use of a PTFE‐coated Jostent, the false aneurysm could be percutaneously closed. However, 6 months later, a new false aneurysm, probably due to focal perforation of the covered Jostent, developed. Cathet. Cardiovasc. Intervent. 50:105–108, 2000.Keywords
This publication has 7 references indexed in Scilit:
- Coronary stent grafts covered by a polytetrafluoroethylene membraneThe American Journal of Cardiology, 1999
- Treatment of a coronary aneurysm with a new polytetrafluoethylene-coated stent: A case reportCatheterization and Cardiovascular Interventions, 1999
- Obliteration of a Coronary Artery Aneurysm With a PTFE-Covered Stent: Endoluminal Graft for Coronary Disease RevisitedCatheterization and Cardiovascular Interventions, 1999
- Successful treatment of false aneurysm of a saphenous vein bypass graft with fistula to the anterior chest wall using “covered” intracoronary stentsHeart, 1998
- Effects of balloon dilatation on ePTFE structural characteristicsJournal of Biomedical Materials Research, 1997
- Coronary artery aneurysm: A reviewProgress in Cardiovascular Diseases, 1997
- Pseudoaneurysm of aortocoronary vein graft secondary to late venous rupture: case report and literature review.Heart, 1993