Vascular graft infections: In vitro and in vivo investigations of a new vascular graft with long‐term protection
- 20 May 2005
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 74B (1) , 601-607
Abstract
We investigated a polyester vascular prosthesis (PET) coated with elemental silver (SC). Measurement of silver release over a period of 52 weeks by means of inductively coupled plasma atomic emission spectrometry of PET with (PET‐G) and without (PET‐N) gelatine impregnation revealed a silver release on the first day of 1.2 ± 0.2μg (PET‐N) and 1.2 ± 0.1 μg (PET‐G) (calculated for 1 g of prosthesis); from the 90th day onward, it was between 0.22 ± 0.14μg (PET‐N) and 0.18 ± 0.12 μg (PET‐G) per day. The prostheses were incubated with Staphylococcus aureus (S.a.), Staphylococcus epidermidis (S.e.), or Escherichia coli (E.c.) to investigate in vitro antibacterial efficacy. After 6 h of incubation, no colony‐forming units were to be seen for any of the bacterial suspensions for PET with SC (p < 0.001). To investigate in vivo antibacterial efficacy, PET‐G rings with and without SC contaminated with S.a., S.e., or E.c. were implanted in 18 albino rabbits and examined 7 days after agar culture for 48 h. The silver coating was associated with a significant reduction in bacterial growth (S.a., p = 0.001; S.e., p < 0.005; E.c., p < 0.001). The silver‐coated prosthesis, with and without gelatine impregnation, had a significantly antibacterial effect with continuous release of silver. © 2005 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2005Keywords
This publication has 43 references indexed in Scilit:
- Comparison of the resistance to infection of rifampin-bonded gelatin-sealed and silver/collagen-coated polyester prosthesesJournal of Vascular Surgery, 2002
- Silver I: its antibacterial properties and mechanism of actionJournal of Wound Care, 2002
- Plaque formation on surface modified dental implantsClinical Oral Implants Research, 2001
- Experimental Treatment of Vascular Graft Infection due to Staphylococcus epidermidis by In Situ Replacement with a Rifampin-bonded Polyester GraftAnnals of Vascular Surgery, 2001
- Anti-Infective Efficacy of Silver-Coated Medical ProsthesesClinical Infectious Diseases, 1999
- Interaction of silver nitrate with readily identifiable groups: relationship to the antibacterialaction of silver ionsLetters in Applied Microbiology, 1997
- The elution and binding characteristics of rifampicin for three commercially available protein-sealed vascular graftsJournal of Antimicrobial Chemotherapy, 1996
- A large randomized clinical trial of a silver-impregnated urinary catheter: Lack of efficacy and staphylococcal superinfectionThe American Journal of Medicine, 1995
- Antibiotic-bonded PTFE vascular grafts: The effect of silver antibiotic on bioactivity following implantationJournal of Surgical Research, 1991
- Control of prosthetic bacterial infection: Evaluation of an easily incorporated, tightly bound, silver antibiotic PTFE graftJournal of Surgical Research, 1988