Efficacy and safety of a self‐expandable stent‐based endobiliary radiofrequency ablation: A preclinical study
Open Access
- 5 November 2026
- journal article
- Published by Wiley in Bioengineering & Translational Medicine
Abstract
The temperature‐controlled, bipolar self‐expandable stent (SES)‐based radiofrequency ablation (RFA) system represents a novel endobiliary RFA approach; however, its efficacy and safety must be verified before clinical application. The study aimed to investigate the efficacy and safety of SES–RFA in a porcine common bile duct (CBD) model. Ablation ranges were investigated using ex vivo porcine liver according to radiofrequency energy and time. For the in vivo study, radiofrequency parameters were set at 20 W for a 20‐mm electrode and 25 W for a 25‐mm electrode at 70°C for 120 s. Twelve pigs underwent endoscopic‐guided SES–RFA using either electrode ( n = 6 each). Animals were randomly euthanized at 24 h or 2 weeks after SES–RFA ( n = 3 per group). The RFA‐treated CBDs were evaluated using cholangiography, cholangioscopy, hematology, histology, and immunohistochemistry. The SES–RFA system was successfully deployed and retrieved through‐the‐scope without device failure. All animals survived until the study endpoint without RFA‐related adverse events, including ductal perforation, contrast leakage, or hematoma. Hepatobiliary levels of alkaline phosphatase, aspartate aminotransferase, alanine aminotransferase, and gamma‐glutamyl transferase were significantly higher at 2 weeks than at 24 h post‐SES–RFA (all p < 0.05). Ablation ranges and degree of mucosal damage, including collagen and TUNEL‐positive deposition, were significantly increased at 2 weeks than at 24 h (all p < 0.05). The temperature‐controlled bipolar SES–RFA system effectively and safely produced circumferential coagulative necrosis without ductal perforation in the porcine CBD. The system may be used for managing unresectable malignant biliary obstruction in anatomically challenging regions where conventional RFA is difficult to use.Keywords
Funding Information
- Korea Medical Device Development Fund (RS‐2023‐00238092)
- Asan Institute for Life Sciences, Asan Medical Center (2026IE0004)
- National Research Foundation of Korea (RS‐2026‐25469139)
- National Research Foundation of Korea (RS‐2026‐25475904)
This publication has 32 references indexed in Scilit:
- Stent-based and catheter-based endobiliary radiofrequency ablation for malignant biliary obstruction: an ex-vivo comparative studyScientific Reports, 2026
- Novel self-expandable stent–based endobiliary radiofrequency ablation for unresectable malignant biliary obstructionGastrointestinal Endoscopy, 2024
- Stent-based electrode for radiofrequency ablation in the rat esophagus: a preliminary studyScientific Reports, 2022
- Feasibility of balloon-based endobiliary radiofrequency ablation under cholangioscopy guidance in a swine modelScientific Reports, 2021
- Efficacy and safety of palliative endobiliary radiofrequency ablation using a novel temperature-controlled catheter for malignant biliary stricture: a single-center prospective randomized phase II TRIALSurgical Endoscopy, 2020
- Intraductal temperature-controlled radiofrequency ablation in malignant hilar obstruction: a preliminary study in animals and initial human experienceEndoscopy International Open, 2019
- Long-term results of temperature-controlled endobiliary radiofrequency ablation in a normal swine modelGastrointestinal Endoscopy, 2018
- Analysis of Endoscopic Radiofrequency Ablation of Biliary Malignant Strictures in Pancreatic Cancer Suggests Potential Survival BenefitDigestive Diseases and Sciences, 2015
- Endobiliary radiofrequency ablation for malignant biliary obstructionHepatobiliary & Pancreatic Diseases International, 2013
- Endoscopically applied radiofrequency ablation appears to be safe in the treatment of malignant biliary obstructionGastrointestinal Endoscopy, 2011