Improved slice selection for R2* mapping during cryoablation with eddy current compensation
Authors:
Open Access
- 25 June 2008
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 28 (1) , 190-198
Abstract
Purpose To improve the slice profile and image quality of R2* mapping in the iceball during cryoablation with ultrashort echo time (UTE) imaging by compensating for eddy currents induced by the selective gradient when half‐pulse radiofrequency (RF) excitation is employed to achieve UTEs. Materials and Methods A method to measure both B0 and linear eddy currents simultaneously is first presented. This is done with a least‐square fitting process on calibration data collected on a phantom. Eddy currents during excitation are compensated by redesigning the RF pulse and the selective gradient accordingly, while that resultant from the readout gradient are compensated for during image reconstruction. In vivo data were obtained continuously during the cryoablation experiments to calculate the R2* values in the iceball and to correlate them with the freezing process. Results Image quality degradation due to eddy currents is significantly reduced with the proposed approaches. R2* maps of iceball throughout the cryoablation experiments were achieved with improved quality. Conclusion The proposed approaches are effective for compensating eddy currents during half‐pulse RF excitation as well as readout. TEs as short as 100 μsec were obtained, allowing R2* maps to be obtained from frozen tissues with improved quality. J. Magn. Reson. Imaging 2008;28:190–198.Keywords
This publication has 14 references indexed in Scilit:
- Consistent non‐cartesian off‐axis MRI quality: Calibrating and removing multiple sources of demodulation phase errorsMagnetic Resonance in Medicine, 2006
- Three‐dimensional radial ultrashort echo‐time imaging with T2 adapted samplingMagnetic Resonance in Medicine, 2006
- Rapid fat‐suppressed isotropic steady‐state free precession imaging using true 3D multiple‐half‐echo projection reconstructionMagnetic Resonance in Medicine, 2005
- Magnetic resonance imaging of the Achilles tendon using ultrashort TE (UTE) pulse sequencesClinical Radiology, 2004
- Centering the projection reconstruction trajectory: Reducing gradient delay errorsMagnetic Resonance in Medicine, 2003
- Temperature mapping of frozen tissue using eddy current compensated half excitation RF pulsesMagnetic Resonance in Medicine, 2001
- Ultra-short echo-time 2D time-of-flight MR angiography using a half-pulse excitationMagnetic Resonance in Medicine, 1999
- Simple Correction Method fork-Space Trajectory Deviations in MRIJournal of Magnetic Resonance, 1998
- Selection of a convolution function for Fourier inversion using gridding (computerised tomography application)IEEE Transactions on Medical Imaging, 1991
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988