Early Clinical Experience with Carotid Artery Stenting for Internal Carotid Artery Stenosis Using the Protcas GW Protection System
Authors:
Open Access
- 1 January 2026
- journal article
- Published by The Japanese Society for Neuroendovascular Therapy in Journal of Neuroendovascular Therapy
- Vol. 20 (1)
Abstract
Objective: Carotid artery stenting (CAS) using a distal balloon protection device (DBPD), which had been widely used in Japan, became difficult to perform after the discontinuation of the PercuSurge GuardWire (Medtronic, Minneapolis, MN, USA) in April 2021. The Protcas GW protection device (SB-KAWASUMI LABORATORIES, Kanagawa, Japan) was subsequently developed as a new DBPD; however, its clinical use has not yet been reported. In this study, we describe our initial clinical experience with the Protcas GW and evaluate its efficacy and safety in real-world practice.Methods: We retrospectively analyzed consecutive patients who underwent CAS using the Protcas GW between April 15, 2025 and August 31, 2025 at a single center in Japan. Patient demographics, characteristics of the stenotic lesions, and procedural details were collected. Clinical outcomes included technical success, device-related adverse events, and the composite of any stroke, myocardial infarction, or death during the periprocedural period.Results: A total of 10 patients were included. The median age was 76 years (interquartile range [IQR], 75–84), and the cohort consisted of 9 men and 1 woman. Eight patients had symptomatic stenosis. The median North American Symptomatic Carotid Endarterectomy Trial (NASCET) stenosis severity was 56.5% (IQR, 52.3–65.3), and unstable plaque was identified in 5 cases. The median plaque length was 18.5 mm (IQR, 15–23.8 mm). CAS using the Protcas GW was successful in all cases. One case required device replacement due to inadequate distal balloon inflation; no other device-related adverse events were observed. No periprocedural events occurred in the composite outcome of any stroke, myocardial infarction, or death. Asymptomatic diffusion-weighted imaging-positive lesions were detected in 3 cases on postoperative MRI, with lesion counts of 1, 1, and 2, respectively.Conclusion: In this initial clinical experience, CAS performed using the Protcas GW device was completed successfully without major device-related adverse events, indicating that the procedure can be performed safely.Keywords
This publication has 14 references indexed in Scilit:
- Real-World Outcomes of Carotid Artery Stenting in Symptomatic and Asymptomatic Patients With Carotid Artery StenosisJACC: Cardiovascular Interventions, 2024
- Safety and efficacy of an open-cell stent and double-balloon protection for unstable plaques: analysis of 184 consecutive carotid artery stentingsJournal of NeuroInterventional Surgery, 2019
- Double balloon protection during carotid artery stenting for vulnerable carotid stenosis reduces the incidence of new brain lesionsActa Neurochirurgica, 2016
- A Manual Carotid Compression Technique to Overcome Difficult Filter Protection Device Retrieval during Carotid Artery StentingJournal of Stroke and Cerebrovascular Diseases, 2015
- 2011 ASA/ACCF/AHA/AANN/AANS/ACR/ASNR/CNS/SAIP/SCAI/SIR/SNIS/SVM/SVS Guideline on the Management of Patients With Extracranial Carotid and Vertebral Artery Disease: Executive SummaryCirculation, 2011
- Risk Factors and Complications Associated with Difficult Retrieval of Embolic Protection Devices in Carotid Artery StentingCardioVascular and Interventional Radiology, 2011
- The Carotid Revascularization Endarterectomy Versus Stenting Trial (CREST)Stroke, 2010
- Magnetic resonance plaque imaging to predict the occurrence of the slow-flow phenomenon in carotid artery stenting proceduresNeuroradiology, 2009
- Slow-Flow Phenomenon During Carotid Artery Intervention With Embolic Protection DevicesJACC, 2005
- Protected Carotid-Artery Stenting versus Endarterectomy in High-Risk PatientsNew England Journal of Medicine, 2004