Corneal topography from spectral optical coherence tomography (sOCT)
Open Access
- 4 November 2011
- journal article
- research article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 2 (12) , 3232-3247
Abstract
We present a method to obtain accurate corneal topography from a spectral optical coherence tomography (sOCT) system. The method includes calibration of the device, compensation of the fan (or field) distortion introduced by the scanning architecture, and image processing analysis for volumetric data extraction, segmentation and fitting. We present examples of three-dimensional (3-D) surface topography measurements on spherical and aspheric lenses, as well as on 10 human corneas in vivo. Results of sOCT surface topography (with and without fan-distortion correction) were compared with non-contact profilometry (taken as reference) on a spherical lens, and with non-contact profilometry and state-of-the art commercial corneal topography instruments on aspheric lenses and on subjects. Corneal elevation maps from all instruments were fitted by quadric surfaces (as well as by tenth-order Zernike polynomials) using custom routines. We found that the discrepancy in the estimated radius of curvature from nominal values in artificial corneas decreased from 4.6% (without fan distortion correction) to 1.6% (after fan distortion correction), and the difference in the asphericity decreased from 130% to 5%. In human corneas, the estimated corneal radius of curvature was not statistically significantly different across instruments. However, a Bland-Altman analysis showed consistent differences in the estimated asphericity and corneal shape between sOCT topographies without fan distortion correction and the rest of the measurements.Keywords
This publication has 54 references indexed in Scilit:
- Corneal topography with high-speed swept source OCT in clinical examinationBiomedical Optics Express, 2011
- Distortions of the posterior surface in optical coherence tomography images of the isolated crystalline lens: effect of the lens index gradientBiomedical Optics Express, 2010
- Three-dimensional reconstruction of the crystalline lens gradient index distribution from OCT imagingOptics Express, 2010
- Experimental evaluation of optimized ablation patterns for laser refractive surgeryOptics Express, 2009
- Ultra high-speed swept source OCT imaging of the anterior segment of human eye at 200 kHz with adjustable imaging rangeOptics Express, 2009
- Anterior segment imaging with Spectral OCT system using a high-speed CMOS cameraOptics Express, 2009
- Keratoconus Diagnosis with Optical Coherence Tomography Pachymetry MappingOphthalmology, 2008
- Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomographyBritish Journal of Ophthalmology, 2007
- Measuring total corneal power before and after laser in situ keratomileusis with high-speed optical coherence tomographyJournal of Cataract & Refractive Surgery, 2006
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986