Ultramicroelectrodes Diffusion Peculiarities Accessed Through Square-Wave Voltammetry

Authors:

Abstract
Despite analytical solutions of diffusion equations for planar and spherical cases predict for current the Cottrell-type reciprocal square root time decay, the generally accepted apparent solution for disk electrodes involves higher exponents of time. Redox probe square-wave voltammetry on both graphite and gold (ultra)microelectrodes displays linear dependence of net peak currents on square root of frequency even at the lower end of the latter. Moreover, net peak current has been found to be proportional to the ultramicroelectrode area. Hence, faster than Cottrell-type time decay has not been confirmed experimentally for small disk electrodes.