Clinical performance of a handheld digital infrared monocular pupillometer for measurement of the dark-adapted pupil diameter
Authors:
- Karl C. BentleyKBKarl C. Bentley
- Aleem I. MughalAMAleem I. Mughal
- Sandra M. BrownSBSandra M. Brown
- 1 February 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 36 (2) , 277-281
Abstract
To compare the accuracy of a handheld infrared digital pupillometer and digital infrared photography for measurement of the dark-adapted pupil diameter. Department of Ophthalmology and Visual Sciences, Texas Tech University Health Sciences Center, Lubbock, Texas, USA. The right horizontal pupil diameter in healthy volunteers was measured using a NeurOptics PLR-200 pupillometer and then videographed using the infrared function of a CyberShot video camera after 2 minutes and 5 minutes dark adaptation at 1 lux ambient illumination. The best still image was extracted from the video file, and the horizontal pupil diameter was determined by comparison against an internal photographic length standard using digital image software. Accommodation and alertness were controlled during testing. The mean horizontal pupil diameter by infrared photography after 2 minutes of dark adaptation by subject age was 7.71 mm for ages 20 to 29 years, 6.80 mm for ages 30 to 39 years, 6.53 mm for ages 40 to 49 years, 5.94 mm for ages 50 to 59 years, and 6.01 mm for ages 60 to 69 years. The mean difference (infrared photography minus pupillometer) was +0.09 mm (range +0.30 to -0.14 mm) at 2 minutes of adaptation and +0.07 mm (range +0.25 to -0.13 mm) at 5 minutes. The pupillometer accurately measured the horizontal pupil diameter at 1 lux, with no measurement more than 0.3 mm different from infrared photography measurements. The pupillometer had a slight negative bias that is unlikely to introduce an error greater than 0.5 mm in clinical measurements.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of 2 multiple-measurement infrared pupillometers to determine scotopic pupil diameterJournal of Cataract & Refractive Surgery, 2006
- Comparison of two pupillometers in determining pupil size for refractive surgeryActa Ophthalmologica Scandinavica, 2006
- Assessment of pupil size under different light intensities using the Procyon pupillometerJournal of Cataract & Refractive Surgery, 2006
- Comparison of Colvard pupillometer and infrared digital photography for measurement of the dark-adapted pupil diameterJournal of Cataract & Refractive Surgery, 2005
- Scotopic measurement of normal pupil size with the Colvard pupillometer and the Nidek auto-refractorContact Lens and Anterior Eye, 2005
- Comparison of dynamic digital pupillometry and static measurements of pupil size in determining scotopic pupil size before refractive surgeryJournal of Cataract & Refractive Surgery, 2005
- Six-month variability of the dark-adapted pupil diameterJournal of Cataract & Refractive Surgery, 2005
- Monocular versus binocular pupillometryJournal of Cataract & Refractive Surgery, 2004
- Correlation of infrared pupillometers and CCD-camera imaging from aberrometry and videokeratography for determining scotopic pupil sizeJournal of Cataract & Refractive Surgery, 2004
- Day to day variability of the dark-adapted pupil diameterJournal of Cataract & Refractive Surgery, 2004