Woven EndoBridge device apposition and compression using Sim&Size virtual simulation correlate with aneurysm occlusion status: a retrospective cohort study
- 16 October 2024
- journal article
- research article
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 17 (11) , 1216-1222
Abstract
Background Virtual simulation is increasingly used for aneurysm treatment. This study aimed to explore whether mechanical behavior biomarkers of the Woven EndoBridge (WEB) device as computed by Sim&Size simulation software were associated with aneurysmal occlusion status at follow-up. Methods Consecutive patients with aneurysms treated with WEB were retrospectively enrolled (January 2014 to December 2021). Aneurysms were included if three-dimensional digital subtraction angiography and follow-up imaging were available. Device apposition and compression within the aneurysm sac were retrospectively calculated by Sim&Size simulation software. Mean global and maximum compression, mid-device and neck compression, and the Spruce index of heterogeneity of deformation of the simulated device were calculated. A multivariate Lasso regression was performed. Results A total of 81 aneurysms in 80 patients (56 females; mean age 60±12 years) were analyzed. At a mean radiological follow-up of 3±2 years, 62 (77%) showed an adequate occlusion. Mean apposition in the inadequate and adequate occlusion group was 42.9±11% and 53.97±12%, respectively (P=0.002), mean global compression was 7.85±6% and 4.95±3%, respectively, and maximal compression was 20.73±11% and 13.44±6%, respectively (P=0.009). Compressions at mid-level and neck level were comparable between the two groups. The Spruce index was higher in the inadequate occlusion group (0.16±0.05 vs 0.20±0.05, P=0.005). Multivariate analysis showed that wall apposition, maximum compression and the Spruce index were independent prognosticators of aneurysm occlusion. Conclusions Wall apposition, maximum compression and the Spruce index computed by Sim&Size software predicted the likelihood of aneurysm occlusion at follow-up, after WEB treatment.Keywords
This publication has 16 references indexed in Scilit:
- Safety and effectiveness of the Woven EndoBridge (WEB) system for the treatment of wide necked bifurcation aneurysms: final 5 year results of the pivotal WEB Intra-saccular Therapy study (WEB-IT)Journal of NeuroInterventional Surgery, 2023
- Aneurysm treatment with the Woven EndoBridge (WEB) device in the combined population of two prospective, multicenter series: 5-year follow-upJournal of NeuroInterventional Surgery, 2022
- Long-term clinical and angiographic outcome of the Woven EndoBridge (WEB) for endovascular treatment of intracranial aneurysmsScientific Reports, 2022
- Determinants of cerebral aneurysm occlusion after embolization with the WEB device: a single-institution series of 215 cases with angiographic follow-upJournal of NeuroInterventional Surgery, 2022
- Comparison of woven endobridge device sizing with conventional measurements and virtual simulation using the sim&size software: a multicenter experienceJournal Of Neuroradiology, 2022
- Effectiveness, Safety and Risk Factors of Woven EndoBridge Device in the Treatment of Wide-Neck Intracranial Aneurysms: Systematic Review and Meta-AnalysisWorld Neurosurgery, 2020
- Virtual simulation with Sim&Size software for Pipeline Flex Embolization: evaluation of the technical and clinical impactJournal of NeuroInterventional Surgery, 2020
- How to WEB: a practical review of methodology for the use of the Woven EndoBridgeJournal of NeuroInterventional Surgery, 2020
- The safety and effectiveness of the Woven EndoBridge (WEB) system for the treatment of wide-necked bifurcation aneurysms: final 12-month results of the pivotal WEB Intrasaccular Therapy (WEB-IT) StudyJournal of NeuroInterventional Surgery, 2019
- Endovascular Treatment of Unruptured Intracranial Aneurysms by the Woven EndoBridge Device (WEB): Are There Any Aspects Influencing Aneurysm Occlusion?World Neurosurgery, 2018