A comprehensive characterization of the temporal characteristics of dose driven continuous scanning in proton therapy
- 7 July 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (14) , 145003
Abstract
Objective. To comprehensively characterize the temporal characteristics of dose-driven continuous scanning (DDCS) proton therapy system including the rise time to nominal beam current ( t rise ) , minimum time between break spot ( t b ) , beam current fluctuation and the flap dose. Approach. Measurements were performed on a Hitachi proton therapy synchrotron. The beam current fluctuation was measured using a 20 kHz 2D strip ionization chamber (CROSSmini) with a fixed point irradiation without break points (BPs). t rise , t b and flap dose were quantified using an oscilloscope connected to the radiofrequency knockout (RFK), high-speed switching magnet (HSST), and dose monitor signals. These measurements were performed with BPs. t rise was measured and compared using both the analytical log file based and direct measurement approaches. The measurements were performed across three energies (70.2, 150.2 and 228.7 MeV), three beam currents (8, 14 and 20 MU s−1) and five different spot MUs (6, 10, 20, 30 and 40 MU). Main results. The instantaneous beam current measured by CROSSmini showed fluctuations of up to 65%. After applying a 1.0 ms moving average, the beam current coincides with the log file recorded beam current, thereby validating the latter data. The time delay between RFK off and HSST on was (0.08 ± 0.01) ms and minimal flap dose was recorded with an average of 4 μMU. The minimum time between break spots was measured to be (1.67 ± 0.04) ms. The analytical log file-based and direct measurement of t rise were correlated with a Pearson correlation coefficient of 0.86. The t rise measurement across the entire spills showed a dependence on beam current where 8 MU s−1 had a notably higher t rise than 14 and 20 MU s−1. Significance. This is the first comprehensive study to characterize the temporal properties of a DDCS proton therapy system. The methodology introduced in this work would help in the proton DDCS commissioning efforts in other centers.Keywords
This publication has 29 references indexed in Scilit:
- Dosimetric study of synchrotron rapid beam off control and skip spot function for high beam intensity proton therapyMedical Physics, 2024
- Beam delivery characteristics of the Hitachi carbon ion scanning system at Osaka Heavy Ion Medical Accelerator in Kansai (HIMAK)Medical Physics, 2023
- The effect of spill change on reliable absolute dosimetry in a synchrotron proton spot scanning systemMedical Physics, 2023
- Past, present and future of proton therapy for head and neck cancerOral Oncology, 2020
- Technical challenges for FLASH proton therapyPhysica Medica, 2020
- Multiple energy extraction reduces beam delivery time for a synchrotron-based proton spot-scanning systemAdvances in Radiation Oncology, 2018
- Empowering Intensity Modulated Proton Therapy Through Physics and Technology: An OverviewInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Proton therapy – Present and futureAdvanced Drug Delivery Reviews, 2017
- Quality of life outcomes in proton and photon treated pediatric brain tumor survivorsRadiotherapy and Oncology, 2014
- Intrafraction Prostate Translations and Rotations During Hypofractionated Robotic Radiation Surgery: Dosimetric Impact of Correction Strategies and MarginsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2014