The feasibility of using a conventional flexible RF coil for an online MR-guided radiotherapy treatment
Authors:
- Stan J HoogcarspelSHStan J Hoogcarspel
- Sjoerd P M CrijnsSCSjoerd P M Crijns
- Jan J W LagendijkJLJan J W Lagendijk
- Marco van VulpenMVMarco van Vulpen
- Bas W RaaymakersBRBas W Raaymakers
- 27 February 2013
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (6) , 1925-1932
Abstract
The purpose of this paper is to evaluate the impact of a flexible radiofrequency coil on the treatment delivery of an online MR-guided radiotherapy treatment. For this study, we used a Synergy MR body coil (Philips, Best) in combination with the current MRL prototype of the UMC Utrecht. The compatibility of the coil is evaluated in two steps. First, we evaluated the dosimetric impact of the MR coil on both a simple and a complex irradiation strategy for treating spinal bone metastases. This tumor site will likely be chosen for the first in-man treatments with the UMC Utrecht MRL system. Second, we investigated the impact of the treatment beam on the MRI performance of the body coil. In case a single posterior-anterior rectangular field was applied, dose to the target volume was underestimated up to 2.2% as a result of beam attenuation in the MR coil. This underestimation however, decreased to 1% when a stereotactic treatment strategy was employed. The presence of the MR coil in or near the distal site of the treatment beam decreased the exit dose when a magnetic field was present. The MRI performance of the coil was unaffected as the result of the radiation. It is feasible to use the Synergy MR body coil for an online MR-guided radiotherapy treatment without any modification to the coil or attenuation correction methods in the planning stage. The effect of the MRI coil on the dose delivery is minimal and there is no effect of the treatment beam on the SNR of the acquired MRI data.Keywords
This publication has 14 references indexed in Scilit:
- Effect of radiation induced current on the quality of MR images in an integrated linac‐MR systemMedical Physics, 2012
- Radiation induced current in the RF coils of integrated linac‐MR systems: The effect of buildup and magnetic fieldMedical Physics, 2012
- Fast online Monte Carlo-based IMRT planning for the MRI linear acceleratorPhysics in Medicine & Biology, 2012
- GPUMCD: A new GPU‐oriented Monte Carlo dose calculation platformMedical Physics, 2011
- Radiation induced currents in MRI RF coils: application to linac/MRI integrationPhysics in Medicine & Biology, 2010
- Real-time fast inverse dose optimization for image guided adaptive radiation therapy–Enhancements to fast inverse dose optimization (FIDO)Journal of Applied Physics, 2009
- First MR images obtained during megavoltage photon irradiation from a prototype integrated linac‐MR systemMedical Physics, 2009
- MRI/linac integrationRadiotherapy and Oncology, 2008
- Flat-panel cone-beam computed tomography for image-guided radiation therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planningMedical Physics, 1998