Comparison of four different screw configurations for the fixation of Fulkerson osteotomy: a finite element analysis
Open Access
- 26 June 2023
- journal article
- research article
- Published by Springer Nature in Journal of Orthopaedics and Traumatology
- Vol. 24 (1) , 1-13
Abstract
Background: Conventionally, two 4.5 mm cortical screws inserted toward the posterior tibial cortex are usually advocated for the fixation of Fulkerson osteotomy. This finite element analysis aimed to compare the biomechanical behavior of four different screw configurations to fix the Fulkerson osteotomy. Materials and methods: Fulkerson osteotomy was modeled using computerized tomography (CT) data of a patient with patellofemoral instability and fixed with four different screw configurations using two 4.5 mm cortical screws in the axial plane. The configurations were as follows: (1) two screws perpendicular to the osteotomy plane, (2) two screws perpendicular to the posterior cortex of the tibia, (3) the upper screw perpendicular to the osteotomy plane, but the lower screw is perpendicular to the posterior cortex of the tibia, and (4) the reverse position of the screw configuration in the third scenario. Gap formation, sliding, displacement, frictional stress, and deformation of the components were calculated and reported. Results: The osteotomy fragment moved superiorly after loading the models with 1654 N patellar tendon traction force. Since the proximal cut is sloped (bevel-cut osteotomy), the osteotomy fragment slid and rested on the upper tibial surface. Afterward, the upper surface of the osteotomy fragment acted as a fulcrum, and the distal part of the fragment began to separate from the tibia while the screws resisted the displacement. The resultant total displacement was 0.319 mm, 0.307 mm, 0.333 mm, and 0.245 mm from the first scenario to the fourth scenario, respectively. The minimum displacement was detected in the fourth scenario (upper screw perpendicular to the osteotomy plane and lower screw perpendicular to the posterior tibial cortex). Maximum frictional stress and maximum pressure between components on both surfaces were highest in the first scenario (both screws perpendicular to the osteotomy plane). Conclusions: A divergent screw configuration in which the upper screw is inserted perpendicular to the osteotomy plane and the lower screw is inserted perpendicular to the posterior tibial cortex might be a better option for the fixation of Fulkerson osteotomy. Level of evidence Level V, mechanism-based reasoning.Keywords
This publication has 40 references indexed in Scilit:
- Titanium as a Biomaterial for Implants2012
- Outcome of surgery for recurrent patellar dislocation based on the distance of the tibial tuberosity to the trochlear grooveThe Journal of Bone and Joint Surgery. British volume, 2010
- The finite element analysis of a fractured tibia applied by composite bone plates considering contact conditions and time-varying properties of curing tissuesComposite Structures, 2010
- Fundamental Biomechanics in Bone Tissue EngineeringSynthesis Lectures on Tissue Engineering, 2010
- A safe zone for the passage of screws through the posterior tibial cortex in tibial tubercle transferThe Knee, 2005
- Prediction of mechanical properties of human trabecular bone by electrical measurementsPhysiological Measurement, 2005
- A Robust Measure of SkewnessJournal of Computational and Graphical Statistics, 2004
- Use of a Modified Elmslie-Trillat Procedure to Improve Abnormal Patellar Congruence AngleThe American Journal of Sports Medicine, 1994
- Strength Pattern of Cancellous Bone at the Ankle JointEngineering in Medicine, 1988
- Anteromedialization of the Tibial Tuberosity for Patellofemoral MalalignmentClinical Orthopaedics and Related Research, 1983