Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system
- 8 March 2019
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 42 (2) , 489-502
Abstract
The accuracy of a computed tomography (CT)-relative electron density (RED) curve may have an indirect impact on the accuracy of dose calculation by a treatment planning system (TPS). This effect has not been previously quantified for input of different CT-RED curves from different CT-scan units in the Monaco TPS. This study aims to evaluate the effect of CT-RED curve on the dose calculation by the Monaco radiotherapy TPS. Four CT images of the CIRS phantom were obtained by different CT scanners. The accuracy of the dose calculation in the three algorithms of the Monaco TPS (Monte Carlo, collapse cone, and pencil beam) is also evaluated based on TECDOC 1583. The CT-RED curves from the CT scanners were transferred to the Monaco TPS to audit the different algorithms of the TPS. The dose values were measured with an ionization chamber in the CIRS phantom. Then, the dose values were calculated by the Monaco algorithms in the corresponding points. For the Monaco TPS and based on TECDOC 1583, the accuracy of the dose calculation in all the three algorithms is within the agreement criteria for most of the points evaluated. For low dose regions, the differences between the calculated and measured dose values are higher than the agreement criteria in a number of points. For the majority of the points, the algorithms underestimate the calculated dose values. It was also found that the use of different CT-RED curves can lead to minor discrepancies in the dose calculation by the Monaco TPS, especially in low dose regions. However, it appears that these differences are not clinically significant in most of the cases.Keywords
Funding Information
- Shahid Beheshti University of Medical Sciences
This publication has 33 references indexed in Scilit:
- Acceptance and Commissioning of a Treatment Planning System Based on Monte Carlo CalculationsTechnology in Cancer Research & Treatment, 2014
- Dosimetric verification of radiotherapy treatment planning systems in Serbia: national auditRadiation Oncology, 2012
- Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: A report of the AAPM TG‐179Medical Physics, 2012
- Patient-Specific 3D Pretreatment and Potential 3D Online Dose Verification of Monte Carlo–Calculated IMRT Prostate Treatment PlansInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Radiation therapy with unflattened photon beams: Dosimetric accuracy of advanced dose calculation algorithmsRadiotherapy and Oncology, 2011
- Accuracy of out-of-field dose calculations by a commercial treatment planning systemPhysics in Medicine & Biology, 2010
- Dosimetric verification of radiotherapy treatment planning systems: Results of IAEA pilot studyRadiotherapy and Oncology, 2008
- Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo‐based photon and electron external beam treatment planningMedical Physics, 2007
- A method for conversion of hounsfield number to electron density and prediction of macroscopic pair production cross-sectionsRadiotherapy and Oncology, 1986
- Factors Affecting the Measurement of Size and CT Number in Computed TomographyInvestigative Radiology, 1981