Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer
- 1 October 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 41 (10) , 2233-2241
Abstract
To assess the precision of ocular biometry with a new noncontact optical low-coherence interferometer (Aladdin) in healthy subjects and patients with cataracts. Eye Hospital of Wenzhou Medical University, Wenzhou, China. Observational cross-sectional study. Eyes from healthy subjects and eyes from patients with cataracts were examined with the new interferometer. Axial length (AL), anterior chamber depth (ACD), keratometry (K), and white-to-white (WTW) values were measured by 2 operators. The test-retest repeatability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability. Different-operator comparison was analyzed with paired t tests and Bland-Altman plots to assess interoperator reproducibility. Ninety-eight people were enrolled for this study, of which 52 eyes were from healthy subjects and 46 eyes were from patients with cataracts. The AL was the most repeatable and reproducible parameter. The ACD, K values, and WTW measurements were highly repeatable in healthy subjects with CoV less than 0.89% and ICC more than 0.94. However, in patients with cataracts, high repeatability could only be observed for AL, ACD, and K values, whereas WTW measurement had test-retest repeatability of 0.80 mm and an ICC of 0.795. Bland-Altman analysis also showed good agreement between the 2 operators for ocular component measurements, except that the WTW in patients with cataracts had wider 95% limits of agreement (range -0.88 to 0.95 mm). The new biometer showed excellent intraoperator repeatability and interoperator reproducibility for AL, ACD, and K values measurements in both groups. The precision of WTW measurements was lower in patients with cataracts. No author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 13 references indexed in Scilit:
- Evaluation of a new optical biometry device for measurements of ocular components and its comparison with IOLMasterBritish Journal of Ophthalmology, 2014
- Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometerJournal of Cataract & Refractive Surgery, 2014
- Validity and repeatability of the Aladdin ocular biometerBritish Journal of Ophthalmology, 2013
- Comparison of the New Biometer OA-1000 with IOLMaster and Tomey AL-3000Current Eye Research, 2013
- A new optical low coherence reflectometry device for ocular biometry in cataract patientsBritish Journal of Ophthalmology, 2009
- Comparison of manual and automated methods to determine horizontal corneal diameterJournal of Cataract & Refractive Surgery, 2004
- Clinical assessment of the Zeiss IOLMasterJournal of Cataract & Refractive Surgery, 2003
- The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™*Ophthalmic and Physiological Optics, 2001
- Clinical results using the Holladay 2 intraocular lens power formulaJournal of Cataract & Refractive Surgery, 2000
- Biometry and Primary Angle-closure Glaucoma among Chinese, White, and Black PopulationsOphthalmology, 1997