Abstract
High-energy photon beams are widely used in radiotherapy owing to their effectiveness in treating deep-seated tumors and their skin-sparing effects. However, treatment planning systems — the most common method for estimating skin dose — have been reported to deviate by as much as 34% from measurements, potentially leading to various side effects. To address this issue, numerous studies have explored skin dosimeters for direct skin dose measurement. However, many of these are limited in clinical applicability because they rely on highly toxic materials such as lead and mercury. In this study, a skin dosimeter based on BaTiO 3 — a nearly non-toxic material — was fabricated, and its fundamental response to high-energy photon beams was evaluated to assess its feasibility for practical applications. The dosimeter was tested for linearity, repeatability, dose-rate dependence, and percent-depth-dose (PDD) conformity. The results demonstrated excellent linearity and repeatability, along with consistent performance across varying dose rates. However, although the dosimeter exhibited excellent D max conformity, the PDD 20/10 values exhibited errors greater than 3.4%, which are likely attributed to several factors, including the atomic number and energy dependence of the sensor. These findings suggest that BaTiO 3 possesses favorable electrical characteristics for use as a skin dosimeter and that a clinically viable device may be achievable through further studies.