Fundamental properties of a perovskite with low dermal toxicity: potential application for skin dosimetry in radiotherapy
- 1 June 2026
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 21 (06) , P06016
Abstract
High-energy photon beams are widely used in radiotherapy owing to their effectiveness in treating deep-seated tumors and their skin-sparing effects. However, treatment planning systems — the most common method for estimating skin dose — have been reported to deviate by as much as 34% from measurements, potentially leading to various side effects. To address this issue, numerous studies have explored skin dosimeters for direct skin dose measurement. However, many of these are limited in clinical applicability because they rely on highly toxic materials such as lead and mercury. In this study, a skin dosimeter based on BaTiO 3 — a nearly non-toxic material — was fabricated, and its fundamental response to high-energy photon beams was evaluated to assess its feasibility for practical applications. The dosimeter was tested for linearity, repeatability, dose-rate dependence, and percent-depth-dose (PDD) conformity. The results demonstrated excellent linearity and repeatability, along with consistent performance across varying dose rates. However, although the dosimeter exhibited excellent D max conformity, the PDD 20/10 values exhibited errors greater than 3.4%, which are likely attributed to several factors, including the atomic number and energy dependence of the sensor. These findings suggest that BaTiO 3 possesses favorable electrical characteristics for use as a skin dosimeter and that a clinically viable device may be achievable through further studies.Keywords
This publication has 21 references indexed in Scilit:
- On dosimetric characteristics of detectors for relative dosimetry in small fields: a multicenter experimental studyPhysics in Medicine & Biology, 2024
- Feasibility study of CdZnTe and CdZnTeSe based high energy X-ray detector using linear acceleratorNuclear Engineering and Technology, 2023
- Development and evaluation of a lead oxide flexible skin dosimeter with a silicon passive layerJournal of Instrumentation, 2022
- Evaluation of two-dimensional dosimeter based on thallium (I) bromide for measurement of surface dose in radiotherapyJournal of Instrumentation, 2021
- Development and evaluation of a monoxide-based flexible skin dosimeter for radiotherapy at photon energiesJournal of Instrumentation, 2021
- Feasibility Evaluation of HgI2-based Flexible Dosimeter for Surface Dose Measurements in RadiotherapyJournal of Instrumentation, 2019
- A simple technique to improve calculated skin dose accuracy in a commercial treatment planning systemJournal of Applied Clinical Medical Physics, 2018
- Development and evaluation of multi-energy PbO dosimeter for quality assurance of image-guide radiation therapy devicesJournal of Instrumentation, 2017
- Reproducibility study of TLD-100 micro-cubes at radiotherapy dose levelApplied Radiation and Isotopes, 1999
- Diode verification of routine electron-beam treatments.Medical Dosimetry, 1999