Abstract
Flexible hybrid electronics (FHE) is a breakthrough in semiconductor packaging and an opportunity in the field on medical applications, such as the monitoring of physiological parameters using wearable devices. Using ultrasound on portable systems is a promising way to probe vascular network and arteries. This article focuses on developing a flexible, thin label that incorporates a piezoelectric micromachined ultrasonic transducer (PMUT) and can be used in a medical patch. The process described is carried out on a silicon wafer carrier in a microelectronic manufacturing line, enabling high integration resolution. PMUTs are hybridized on a flexible interconnection redistribution layer using flip-chip technology; they are thinned down and encapsulated with polyimide film. Finally, the label is attached to a ribbon and fully encapsulated with Parylene that provides biocompatibility and moisture tolerance. The functionality of the flexible PMUT arrays is validated through acoustic measurements in water.