Introduction of new industry standards for cardiac implantable electronic devices: balancing benefits and unexpected risks
Authors:
- 1 March 2012
- journal article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 14 (8) , 1081-1086
Abstract
The DF-4 connector is a novel industry standard for the connection of a defibrillator lead to the generator. It aims at reducing the bulk created by two or three pins at the proximal end of the defibrillator lead and its corresponding ports at the header of the device. Having only one connection port between the lead and the device reduces the material in the pocket, the risk of lead-to-port mismatch, may lower the risk of lead abrasion, and probably makes the implantation procedure a little easier since only one set screw is required. However, all these conceived benefits are related to convenience rather than to a medical need. After the recent experiences with the possible negative clinical impact of 'minor' changes like simply downsizing a defibrillator lead, a word of caution is warranted. The lead is the weakest part of the defibrillator system, complex in design and undergoing constant stress through movement. It is very hard to predict which issues may evolve over time with the changes in lead design. Does the perceived benefit really outweigh an unpredictable risk in a sensitive medical product like a defibrillator? This article tries to address the possible issues of the new spring contacts instead of set screws, the proximity of the low- and high-voltage connections as well as the inability of adding a pace/sense or an additional shock lead without a special adaptor, and advocates a measured speed in the introduction of this technology.Keywords
This publication has 11 references indexed in Scilit:
- Insulation Defects of Thin High-Voltage ICD Leads:Journal of Cardiovascular Electrophysiology, 2011
- Limitations of the DF-4 Defibrillator Connector Necessitating Device RemovalPacing and Clinical Electrophysiology, 2010
- Ventricular oversensing due to manufacturer-related differences in implantable cardioverter-defibrillator signal processing and sensing lead propertiesEP Europace, 2010
- Recommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and GuidelinesHeart Rhythm, 2009
- Testing New ICD TechnologyNew England Journal of Medicine, 2008
- Implantable cardioverter-defibrillator lead disintegration at the level of the tricuspid valveHeart Rhythm, 2008
- Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillatorHeart Rhythm, 2007
- Complications of Implantable Cardioverter Defibrillator Therapy in 440 Consecutive PatientsPacing and Clinical Electrophysiology, 2005
- Pacemaker leads: performance and progressThe American Journal of Cardiology, 1999
- A Voluntary Standard for 3.2 mm Unipolar and Bipolar Pacemaker Leads and ConnectorsPacing and Clinical Electrophysiology, 1986