ASME B89.4.23 Performance Evaluation Tests and Geometry Errors in X-Ray Computed Tomography Systems
Open Access
- 20 January 2022
- journal article
- research article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 126
Abstract
A documentary standard produced by the American Society of Mechanical Engineers (ASME) for performance evaluation of industrial X-ray computed tomography (XCT) systems for dimensional measurements was released in early 2021. This standard, ASME B89.4.23-2020, specifies test procedures that may be performed to determine whether a system meets the manufacturer’s accuracy specifications for acceptance before or after purchase, or for periodic reverification. While there are some core testing requirements in the standard, there is also some flexibility, allowing for a variety of testing configurations that meet the requirements of the standard. It is important that the chosen testing configuration be sensitive to the different systematic sources of error in XCT systems to provide confidence that the system will meet the manufacturer’s accuracy specifications for measurements performed by the user subsequent to testing. In this paper, we provide guidance on how to optimally apply the ASME 89.4.23 standard in industry to achieve high sensitivity to geometry errors in cone-beam XCT systems. Through simulation studies, we present some examples of testing configurations that meet the requirements of the ASME B89.4.23 standard and discuss their sensitivity to geometry errors of the detector and the rotation stage. We show that there are some testing configurations that achieve maximal sensitivity to these errors, while other configurations do not capture these error sources with adequate sensitivity.Keywords
Funding Information
- Physical Measurement Laboratory
This publication has 26 references indexed in Scilit:
- Evaluating the effects of detector angular misalignments on simulated computed tomography dataPrecision Engineering, 2016
- X-ray computed tomography for additive manufacturing: a reviewMeasurement Science and Technology, 2016
- Development of Dimensional X-Ray Computed TomographyInternational Journal of Automation Technology, 2015
- Towards geometrical calibration of x-ray computed tomography systems—a reviewMeasurement Science and Technology, 2015
- A new method of performance verification for x-ray computed tomography measurementsMeasurement Science and Technology, 2014
- Industrial applications of computed tomographyCIRP Annals, 2014
- Physical characterization and performance evaluation of an x-ray micro-computed tomography system for dimensional metrology applicationsMeasurement Science and Technology, 2012
- Accuracy of industrial computed tomography measurements: Experimental results from an international comparisonCIRP Annals, 2012
- Analysis of the effect of cone-beam geometry and test object configuration on the measurement accuracy of a computed tomography scanner used for dimensional measurementMeasurement Science and Technology, 2011
- Computed tomography for dimensional metrologyCIRP Annals, 2011