Optia® continuous mononuclear collection (CMNC) system is a safe and efficient system for hematopoietic progenitor cells‐apheresis (HPC‐a) collection and yields a lower product hematocrit (HCT%) than the COBE® spectra system: A retrospective study
Authors:
- 30 March 2018
- journal article
- research article
- Published by Wiley in Journal of Clinical Apheresis
- Vol. 33 (4) , 505-513
Abstract
Purpose: We evaluated the Optia® continuous mononuclear collection (CMNC) system for hematopoietic progenitor cell‐apheresis (HPC‐A) collection (Terumo BCT, Lakewood, CO) compared to the COBE® Spectra (Terumo BCT, Lakewood, CO), including both large volume leukapheresis (LVL) and non‐LVL collections.Methods: We performed a retrospective data analysis of all autologous HPC‐A collections with the Optia® CMNC system (n = 93; LVL = 59, non‐LVL = 34) since implementation at our institution and compared it with a similar number of concurrent collections utilizing the COBE® Spectra (n = 96; LVL = 68, non‐LVL = 28). The population studied included multiple myeloma (62 patients/171 collections) and lymphoma (5 patients/18 collections). Mobilization was achieved using chemotherapy + G‐CSF (n = 108), chemotherapy + G‐CSF + plerixafor (n = 67), G‐CSF alone (n = 10), or G‐CSF + plerixafor (n = 4). Based on our minimum predicted collection formula and the collection goal, 7‐30 L of whole blood was processed. Per protocol, a minimum of 2 days of collection was performed.Results: HPC‐A collected on Optia® CMNC had lower %HCT than those collected on COBE® Spectra (3.7 versus 4.3%, P = .029). There were no statistically significant differences between the two devices for other variables examined, including preapheresis WBC count and CD34+ cell count, procedure time, whole blood volume processed, collection efficiency (CE2), % platelet loss and throughput. CE2 for both devices was higher when <30 L of whole blood volume was processed. A linear correlation was noted between the preapheresis CD34+ cell count and CD34+ cells collected. No adverse events or bleeding episodes were noted, even when acetyl salicyclic acid (ASA) was given.Conclusions: Optia® CMNC system is equivalent to the COBE® Spectra, with significantly lower product HCT%.Keywords
This publication has 22 references indexed in Scilit:
- Apheresis activity in Spain: A survey of the Spanish Apheresis GroupTransfusion and Apheresis Science, 2013
- Peripheral blood stem cell collection in low‐weight children: retrospective comparison of two apheresis devicesTransfusion, 2013
- Allogeneic donor peripheral blood “stem cell” apheresis: prospective comparison of two apheresis systemsTransfusion, 2011
- Mobilized allogeneic peripheral stem/progenitor cell apheresis with Spectra Optia v.5·0, a novel, automatic interface‐controlled apheresis system: results from the first feasibility trialVox Sanguinis, 2011
- Automatic interface‐controlled apheresis collection of stem/progenitor cells: results from an autologous donor validation trial of a novel stem cell apheresis deviceTransfusion, 2010
- Apheresis Technologies and Clinical Applications: The 2007 International Apheresis RegistryTherapeutic Apheresis and Dialysis, 2010
- Hematopoietic Stem-Cell TransplantationThe New England journal of medicine, 2006
- PBPC collection techniques: standard versus large volume leukapheresis (LVL) in donors and in patientsTransfusion and Apheresis Science, 2005
- Large‐volume leukapheresis yields more viable CD34+ cells and colony‐forming units than normal‐volume leukapheresis, especially in patients who mobilize low numbers of CD34+ cellsTransfusion, 2005
- The ISHAGE Guidelines for CD34+ Cell Determination by Flow CytometryJournal of Hematotherapy, 1996