Controlled saturation magnetization transfer for reproducible multivendor variable flip angle T1 and T2 mapping
Open Access
- 17 December 2019
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 84 (1) , 221-236
Abstract
Purpose: The widespread clinical application of quantitative MRI has been hindered by a lack of reproducibility across sites and vendors. Previous work has attributed this to incorrect B1 mapping or insufficient spoiling conditions. We recently proposed the controlled saturation magnetization transfer (CSMT) framework and hypothesized that the lack of reproducibility can also be attributed to magnetization transfer effects. This work seeks to validate this hypothesis and demonstrate that reproducible multivendor single‐pool relaxometry can be achieved with the CSMT approach.Methods: Three healthy volunteers were scanned on scanners from 3 vendors (GE Healthcare, Philips, Siemens). An extensive set of images necessary for joint T1 and T2 estimation were acquired with (1) each vendor default RF pulses and spoiling conditions; (2) harmonized RF spoiling; and (3) harmonized RF spoiling and CSMT pulses. Different subsets of images were used to generate 6 different T1 and T2 maps for each subject’s data from each vendor. Cross‐protocol, cross‐vendor, and test/retest variability were estimated.Results: Harmonized RF spoiling conditions are insufficient to ensure good cross‐vendor reproducibility. Controlled saturation magnetization transfer allows cross‐protocol variability to be reduced from 18.3% to 4.0%. Whole‐brain variability using the same protocol was reduced from a maximum of 19% to 4.5% across sites. Both CSMT and native vendor RF conditions have a reported variability of less than 5% for repeat measures on the same vendor.Conclusion: Magnetization transfer effects are a major contributor to intersite/intrasite variability of T1 and T2 estimation. Controlled saturation magnetization transfer stabilizes these effects, paving the way for the use of single‐pool T1 and T2 as a reliable source for clinical diagnosis across sites.Keywords
Funding Information
- European Research Council (FP7/20072013/ERC, [319456] dHCP project)
- National Institute for Health Research
- Wellcome Trust (WT 203148/Z/16/Z)
- Engineering and Physical Sciences Research Council (EP/L015226/1)
This publication has 31 references indexed in Scilit:
- Quantitative multi-parameter mapping of R1, PD*, MT, and R2* at 3T: a multi-center validationFrontiers in Neuroscience, 2013
- Multi-site voxel-based morphometry — Not quite there yetNeuroImage, 2011
- Unified segmentation based correction of R1 brain maps for RF transmit field inhomogeneities (UNICORT)NeuroImage, 2011
- Rapid B1+ mapping using a preconditioning RF pulse with TurboFLASH readoutMagnetic Resonance in Medicine, 2010
- B1 mapping by Bloch‐Siegert shiftMagnetic Resonance in Medicine, 2010
- Multisite neuroimaging trialsCurrent Opinion in Neurology, 2009
- Standardized structural magnetic resonance imaging in multicentre studies using quantitative T 1 and T 2 imaging at 1.5 TNeuroImage, 2008
- Actual flip‐angle imaging in the pulsed steady state: A method for rapid three‐dimensional mapping of the transmitted radiofrequency fieldMagnetic Resonance in Medicine, 2006
- Improved Optimization for the Robust and Accurate Linear Registration and Motion Correction of Brain ImagesNeuroImage, 2002
- Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithmIEEE Transactions on Medical Imaging, 2001