Accuracy and reliability of a new, protractor-based neck goniometer
Authors:
- Michael John HaynesMHMichael John Haynes
- Stephen EdmondstonSEStephen Edmondston
- 1 November 2002
- journal article
- clinical trial
- Published by National University of Health Sciences in Journal of Manipulative and Physiological Therapeutics
- Vol. 25 (9) , 579-586
Abstract
To assess the reliability of the SpinT, a new protractor-based device, for measuring active cervical spine ranges of motion. In addition, to compare the accuracy of the Cervical Ranges of Motion (CROM) instrument and SpinT measurements of rotation about the Y axis with and without tilt, the former motion occurring during natural rotation of the head. Interexaminer reliability, intraexaminer reliability, and accuracy trials were conducted. Two examiners made 2 individual measurements of each of the individual cervical ranges of motion of 23 patients (15 men, 8 women; aged 21 to 42 years) with no cervical symptoms. The patients were asked to move their necks to end range while they sat upright. The accuracy of the CROM instrument and SpinT goniometers was assessed with a testing instrument capable of rotating and/or tilting to preset angles and upon which either device could be positioned. There was excellent agreement between the SpinT measurements of rotation about the Y axis compared with the readings from the testing platform regardless of the angle of tilt, whereas the CROM instrument displayed poor concordance when the tilt exceeded 5 degrees. The reliability trials generally yielded close agreement between the examiners, especially regarding measurements of rotation left and right and extension and revealed higher concordance regarding intraexaminer results. This study indicates that SpinT measurements of active cervical ranges of motion are reliable and that the SpinT goniometer accurately measures rotation with associated tilt.Keywords
This publication has 10 references indexed in Scilit:
- Reproducibility and instrument validity of a new ultrasonography-based system for measuring cervical spine kinematicsClinical Biomechanics, 2000
- Assessment of published reliability studies for cervical spine range-of-motion measurement toolsJournal of Manipulative and Physiological Therapeutics, 2000
- Precision and accuracy of an electrogoniometerJournal of Manipulative and Physiological Therapeutics, 1999
- Three-dimensional analysis of active cervical motion: the effect of age and genderClinical Biomechanics, 1996
- Forming inferences about some intraclass correlation coefficients.Psychological Methods, 1996
- Statistical Methodology for the Concurrent Assessment of Interrater and Intrarater Reliability: Using Goniometric Measurements as an ExamplePhysical Therapy, 1994
- Age and Gender Related Normal Motion of the Cervical SpineSpine, 1992
- Intertester Reliability of the Cervical Range of Motion DeviceJournal of Orthopaedic & Sports Physical Therapy, 1992
- Reliability of Measurements of Cervical Spine Range of Motion—Comparison of Three MethodsPhysical Therapy, 1991
- Three-Dimensional Analysis of Neck MotionSpine, 1990