Accuracy and Equivalence of Particle Number Concentration Measurements (0.3–10 µm) from a Low-Cost Sensirion SPS30 Compared with the OPS 3330 Under Field Conditions
Open Access
- 8 August 2026
- journal article
- research article
- Published by MDPI AG in Sustainability
- Vol. 18 (16) , 8097
Abstract
Mass concentrations of particulate matter are a fundamental metric for air quality and health impact assessment; however, they are insufficient for accurately characterizing exposure. They do not capture particle size distribution or number concentration. Therefore, aerosol assessment should include particle number concentration (PNC), which better represents toxicologically relevant fractions and enables more precise source identification. The aim of this study was to conduct a comprehensive evaluation of particle number concentration (PNC) measurements in the 0.3–10 µm size range obtained using three low-cost Sensirion SPS30 particle sensors under field conditions in an urban environment. The analyses included an assessment of agreement between the SPS30 sensors, an evaluation of their measurement performance against the OPS 3330 optical particle spectrometer, and the development of calibration models. The SPS30 sensors showed high inter-device repeatability for PNC in the 0.3–1.0 µm range (CVd < 2%). However, measurement performance declined with increasing particle size, with the index of agreement (IOA) decreasing from 0.8 (0.3–0.5 µm) to −0.5 (2.5–10 µm). Sensor accuracy was influenced by meteorological conditions: relative humidity primarily affected short-term variability (precision and dynamic agreement), while temperature controlled systematic bias. Although incorporating these variables into advanced calibration models improved performance, SPS30 sensors remained unsuitable for PNC measurements in the 2.5–10 µm range, exhibiting systematic errors of ~25% even after nonlinear correction. The findings support the responsible use of low-cost particle sensors for supplementary air quality monitoring, contributing to accessible environmental data and sustainable urban air quality management.Keywords
Funding Information
- Ministry of Science and Higher Education of the Russian Federation (150000-501.00-10000)
This publication has 72 references indexed in Scilit:
- Quantifying road traffic influence on coarse and accumulation mode particle number concentration in a street canyonEnvironmental Science and Pollution Research, 2026
- Study on calibration of low-cost particulate matter sensors for hydrophilic and hydrophobic particles under varying relative humidityAir Quality, Atmosphere & Health, 2025
- Performance Evaluation of Five Different Low-Cost Particulate Matter Sensors for Monodisperse Test AerosolsAerosol and Air Quality Research, 2025
- Performance and data acquisition from low-cost air quality sensors: a comprehensive reviewAir Quality, Atmosphere & Health, 2024
- In chamber calibration and performance evaluation of air quality low-cost sensorsAtmospheric Pollution Research, 2024
- The influence of COVID-19 pandemic on deposited dose of outdoor particulate matter in human respiratory tract: A case study from Krakow, Southern PolandAir Quality, Atmosphere & Health, 2024
- Assessment of the Kraków inhabitants’ health risk caused by the exposure to inhalation of outdoor air contaminantsStochastic Environmental Research and Risk Assessment, 2016
- Heavy metals from non-exhaust vehicle emissions in urban and motorway road dustsEnvironmental Monitoring and Assessment, 2016
- Explaining the high PM10 concentrations observed in Polish urban areasAir Quality, Atmosphere & Health, 2015
- High-PM10 concentration episodes in Seoul, Korea: Background sources and related meteorological conditionsAtmospheric Environment, 2011