UV-C Light-Based Surface Disinfection: Analysis of Its Virucidal Efficacy Using a Bacteriophage Model
Open Access
- 10 March 2022
- journal article
- research article
- Published by MDPI AG in International Journal of Environmental Research and Public Health
- Vol. 19 (6) , 3246
Abstract
Background: The reprocessing of medical devices has become more complex due to increasing hygiene requirements. Previous studies showed satisfactory bactericidal disinfection effects of UV-C light in rigid and flexible endoscopes. Especially in the context of the current COVID-19 pandemic, virucidal properties are of high importance. In the present study, the virucidal efficacy of UV-C light surface disinfection was analyzed. Methods: MS-2 bacteriophages were applied to the test samples and irradiated by UV-C light using the UV Smart D25 device; unirradiated test samples were used as controls. A dilution series of the samples was mixed with Escherichia coli and assayed. Results: 8.6 × 1012 pfu could be harvested from the unprocessed test samples. In the control group without UV-C exposure, a remaining contamination of 1.2 × 1012 pfu was detected, resulting in a procedural baseline reduction rate with a LOG10 reduction factor of 0.72. The LOG10 reduction factor was found to be 3.0 after 25 s of UV-C light exposure. After 50 and 75 s of UV-C radiation LOG10 reduction factors 4.2 and 5.9, respectively, were found, with all reductions being statistically significantly different to baseline. Conclusions: The tested UV system seems to provide a significant virucidal effect after a relatively short irradiation time.
Keywords
Funding Information
- UV Smart Technologies B.V. (001)
This publication has 26 references indexed in Scilit:
- UV emissions from low energy artificial light sourcesPhotodermatology, Photoimmunology & Photomedicine, 2014
- Anforderungen an die Hygiene bei der Aufbereitung von MedizinproduktenBundesgesundheitsblatt - Gesundheitsforschung - Gesundheitsschutz, 2012
- Human Papillomavirus and Rising Oropharyngeal Cancer Incidence in the United StatesJournal of Clinical Oncology, 2011
- Terminal Decontamination of Patient Rooms Using an Automated Mobile UV Light UnitInfection Control & Hospital Epidemiology, 2011
- Inactivation of murine norovirus, feline calicivirus and echovirus 12 as surrogates for human norovirus (NoV) and coliphage (F+) MS2 by ultraviolet light (254 nm) and the effect of cell association on UV inactivationLetters in Applied Microbiology, 2011
- RNA under attack: Cellular handling of RNA damageCritical Reviews in Biochemistry and Molecular Biology, 2009
- Surface disinfection by exposure to germicidal UV lightIndian Journal of Medical Microbiology, 2008
- High-Level Disinfection or “Sterilization” of Endoscopes?Infection Control & Hospital Epidemiology, 1996
- Ultraviolet inactivation of selected bacteria and viruses with photoreactivation of the bacteriaWater Research, 1987
- Anticipatics, handwashing, and handwashing facililtiesAmerican Journal of Infection Control, 1983