Abstract
Ion beam focusing for a nuclear microprobe is typically achieved by an operator adjusting the currents supplied to a series of quadrupole magnets, while observing the resulting changes in size and shape of the beam spot via an optical microscope image. This process requires an experienced operator to implement in a reasonable time frame. The present work demonstrates an automated, software-based solution to ion beam focusing. The software determines parameters describing the size and shape of the beam spot in real time, by processing images captured from the microscope. These parameters then provide feedback to a minimisation routine based on the downhill simplex algorithm, which controls the currents supplied to the quadrupole focusing magnets. The minimal beam spot size is achieved in less than three minutes, regardless of the experience of the human operator. The physical microprobe system, on which the software is demonstrated, is not presently capable of achieving spot sizes much below 3 µm, and thus the performance of the software below this limit is still to be determined. Within the limits tested, however, the software demonstrates consistent performance, and greatly enhances the accessibility of the microprobe for less experienced users.
Funding Information
  • Vetenskapsrådet (VR-RFI-2019-00191)
  • Uppsala University

This publication has 28 references indexed in Scilit: