Adding customized electron energy beams to TrueBeam linear accelerators
Authors:
Open Access
- 9 May 2022
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (7) , e13633
Abstract
Purpose: To better meet clinical needs and facilitate optimal treatment planning, we added two new electron energy beams (7 and 11 MeV) to two Varian TrueBeam linacs.Methods: We worked with the vendor to create two additional customized electron energies without hardware modifications. For each beam, we set the bending magnet current and then optimized other beam‐specific parameters to achieve depths of 50% ionization (I50) of 2.9 cm for 7 MeV and 4.2 cm for the 11 MeV beam with the 15 × 15 cm2 cone at 100 cm source‐to‐surface distance (SSD) by using an ionization chamber profiler (ICP) with a double‐wedge (DW) phantom. Beams were steered and balanced to optimize symmetry with the ICP. After all parameters were set, full commissioning was done including measuring beam profiles, percent depth doses (PDDs), output factors (OFs) at standard, and extended SSDs. Measured data were compared between the two linacs and against the values calculated by our RayStation treatment planning system (TPS) following Medical Physics Practice Guideline 5.a (MPPG 5.a) guidelines.Results: The I50 values initially determined with the ICP/DW agreed with those from a PDD‐scanned in‐water phantom within 0.2 mm for the 7 and 11 MeV on both linacs. Comparison of the beam characteristics from the two linacs indicated that flatness and symmetry agreed within 0.4%, and point‐by‐point differences in PDD were within 0.01% ± 0.3% for the 7 MeV and 0.01% ± 0.3% for the 11 MeV. The OF ratios between the two linacs were 1.000 ± 0.007 for the 7 MeV and 1.004 ± 0.007 for the 11 MeV. Agreement between TPS‐calculated outputs and measurements were −0.1% ± 1.0% for the 7 MeV and 0.2% ± 0.8% for the 11 MeV. All other parameters met the MPPG 5.a's 3%/3‐mm criteria.Conclusion: We were able to add two new beam energies with no hardware modifications. Tuning of the new beams was facilitated by the ICP/DW system allowing us to have the procedures done in a few hours and achieve highly consistent results across two linacs.PACS numbers: 87.55.Qr, 87.56.FcKeywords
This publication has 7 references indexed in Scilit:
- Monitoring linear accelerators electron beam energy constancy with a 2D ionization chamber array and double‐wedge phantomJournal of Applied Clinical Medical Physics, 2019
- Quantification of beam steering with an ionization chamber arrayJournal of Applied Clinical Medical Physics, 2018
- AAPM Medical Physics Practice Guideline 5.a.: Commissioning and QA of Treatment Planning Dose Calculations — Megavoltage Photon and Electron BeamsJournal of Applied Clinical Medical Physics, 2015
- Clinical implementation of electron energy changes of Varian linear acceleratorsJournal of Applied Clinical Medical Physics, 2009
- Field equivalence for clinical electron beamsPhysics in Medicine & Biology, 2000
- Optimum field size and choice of isodose lines in electron beam treatmentInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995
- Effect of Air Space on Depth Dose in Electron Beam TherapyRadiology, 1978