Clinical implementation of electron energy changes of Varian linear accelerators
Authors:
Open Access
- 1 September 2009
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 10 (4) , 177-187
Abstract
Modern dual photon energy linear accelerators often come with a few megavoltage electron beams. The megavoltage electron beam has limited range and relative sharp distal falloff in its depth dose curve compared to that of megavoltage photon beam. Its radiation dose is often delivered appositionally to cover the target volume to its distal 90% depth dose (d90), while avoiding the normal ‐ sometimes critical ‐ structure immediately distal to the target. Varian linear accelerators currently offer selected electron beams of 4, 6, 9, 12, 16 and 20 MeV electron beam energies. However, intermediate electron energy is often needed for optimal dose distribution. In this study we investigated electron beam characteristics and implemented two intermediate 7 and 11 MeV electron beams on Varian linear accelerators. Comprehensive tests and measurements indicated the new electron beams met all dosimetry parameter criteria and operational safety standards. Between the two new electron beams and the existing electron beams we were able to provide a choice of electron beams of 4, 6, 7, 9, 11, 12, 16 and 20 MeV electron energies, which had d90 depth between 1.5 cm and 6.0 cm (from 1.5 cm to 4.0 cm in 0.5 cm increments) to meet our clinical needs.PACS number: 87.56.bd.Keywords
This publication has 9 references indexed in Scilit:
- A beam-matching concept for medical linear acceleratorsActa Oncologica, 2009
- Accelerator beam data commissioning equipment and procedures: Report of the TG‐106 of the Therapy Physics Committee of the AAPMMedical Physics, 2008
- Comparison of electron beam characteristics from multiple acceleratorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Comparison of characteristics of photon and electron beams generated by Philips/Elekta and Varian linear acceleratorsMedical Physics, 2002
- Comparative measurements on a series of accelerators by the same vendorMedical Physics, 1999
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Characteristic parameters of 6-21 MeV electron beams from a 21 MeV linear acceleratorMedical Physics, 1991
- Clinical electron‐beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25Medical Physics, 1991
- Evaluation of a photon and an electron beam of a 6-MV linear acceleratorMedical Physics, 1988