Left Ventricular endocardial pacing by the interventricular septum route

Authors:

Abstract
A 75 years old female patient with heart failure (HF) had a cardiac resynchronisation therapy pacemaker device implanted but developed untreatable phrenic nerve stimulation and recurrence of HF after disabling left ventricular (LV) pacing. She agreed with implantation of an LV endocardial lead using an approach through the interventricular septum. A coronary angiogram was performed to be informed about the size and position of septal branches. A deflectable 12F St Jude Medical Agilis 408310 sheath was advanced into the right ventricle towards the interventricular septum. The dilator of the Agilis sheath was advanced to the septum but penetrated without applying noticeable force. The position was confirmed by fluoroscopy, intra cardiac echo, and LV cavity pressure. (left panel) The Agilis sheath was changed for a Medtronic Attain deflectable coronary sinus catheter through which a Medtronic SelectSecure 3830 was implanted in the LV posterolateral area. After checking pacing and sensing parameters, lead stability and proper fixation, the Attain catheter was slit and removed. The lead was provided with ample slack in the LV (right panel). After implant, the patient developed pericardial effusion which was relieved by pericardiocentesis. Perforation of the LV lead was excluded by computed tomography scan and electrical parameters. She recovered completely free of HF symptoms.