A simple computational method is proposed for estimating the time-dependent flux of an x-ray spectrum over domain from data obtained by an -channel array of filtered detectors. It is assumed that the data are related to the spectrum by a discrete, inhomogeneous, first-kind Fredholm integral equation , where is the known response function for each detector channel of the diagnostic. The proposed method constructs a spectral sensitivity for the diagnostic array as a linear combination of the responses, where the coefficients are obtained by a least-squares criterion plus a constraint. The values, once determined, apply as long as the responses are valid. The flux estimate is then simply , without a spectral unfold of the data. The method is useful for quick analyses of time-dependent data, for comparisons with other flux-measuring diagnostics, and for the experimental design of filtered-detector arrays. The method is applied to a five-channel array of filtered photoemissive x-ray detectors [G. A. Chandler et al., Rev. Sci. Instrum. 70, 561 (1999)], used for -pinch measurements at the Z-accelerator facility [R. B. Spielman et al., Phys. Plasmas 5, 2105 (1998)]. Comparisons with unfold results are made, and a first-order analysis of error propagation into is presented.
Keywords
This publication has 14 references indexed in Scilit:
- X-ray emission frompinches atA: Current scaling, gap closure, and shot-to-shot fluctuationsPhysical Review E, 2004
- DMX: An absolutely calibrated time-resolved broadband soft x-ray spectrometer designed for MJ class laser-produced plasmas (invited)Review of Scientific Instruments, 2001
- Diagnostics on Z (invited)Review of Scientific Instruments, 2001
- Spectral resolution for a five-element, filtered, x-ray detector array using the method of Backus and GilbertReview of Scientific Instruments, 2000
- Filtered x-ray diode diagnostics fielded on the Z accelerator for source power measurementsReview of Scientific Instruments, 1999
- Fast resistive bolometryReview of Scientific Instruments, 1999
- Tungsten wire-array Z-pinch experiments at 200 TW and 2 MJPhysics of Plasmas, 1998
- A simple unfold method for Rutherford scattering, intense ion-beam spectrographsReview of Scientific Instruments, 1992
- Measurement of 0.1–3-keV x rays from laser plasmasReview of Scientific Instruments, 1986
- Flat-response, subkiloelectronvolt x-ray detector with a subnanosecond time responseReview of Scientific Instruments, 1978