Anatomically informed GREIT reconstruction: improving EIT imaging for lung monitoring
Open Access
- 5 February 2026
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 47 (2)
Abstract
Objective: Time-difference electrical impedance tomography (EIT) is gaining widespread use for bedside lung monitoring in intensive care patients suffering from lung-related diseases. It involves collecting voltage measurements from electrodes placed on the patient’s thorax, which are then used to reconstruct impedance images. This study investigates how incorporating anatomical information from CT data into the widely used GREIT reconstruction algorithm affects EIT images and improves their interpretability. Approach: Based on clinically motivated lung state scenarios, we simulated EIT measurements to assess how the GREIT parameters influence the result of EIT image reconstruction, particularly with respect to noise performance and image accuracy. We introduce quality measures that allow us to perform a quantitative assessment of reconstruction quality. We incorporate the anatomical features of a patient from CT data by customizing the background conductivity and the distribution of GREIT training targets. Main results: Our analysis confirmed that unphysiological background conductivity assumptions can lead to misleading EIT images, whereas physiological values, although more accurate, come with higher noise sensitivity. By increasing the number of GREIT training targets inside the lung and adapting the respective weighting radius, we significantly improved the anatomical accuracy of the EIT images. When applied to clinical EIT data from a representative ARDS patient, these adjustments in the reconstruction setup substantially enhanced the interpretability of the resulting EIT images. Significance: Incorporating CT-based anatomical data in the GREIT reconstruction significantly enhances the clinical applicability of EIT in lung monitoring. The improved interpretability of EIT images facilitates better-informed clinical decisions and the individualized adjustment of ventilation strategies for critically ill patients.Keywords
All Related Versions
Funding Information
- Deutsche Forschungsgemeinschaft (BE6526/1-1)
- Deutsche Forschungsgemeinschaft (WA1521/26-1)
- HORIZON EUROPE European Research Council (101021526-BREATHE)
This publication has 29 references indexed in Scilit:
- Automatic multiorgan segmentation in thorax CT images using U‐net‐GANMedical Physics, 2019
- Electrical impedance tomography in adult patients undergoing mechanical ventilation: A systematic reviewJournal of Critical Care, 2016
- 3D EIT image reconstruction with GREITPhysiological Measurement, 2016
- Tracking boundary movement and exterior shape modelling in lung EIT imagingPhysiological Measurement, 2015
- The influence of image reconstruction algorithms on linear thorax EIT image analysis of ventilationPhysiological Measurement, 2014
- Uniform background assumption produces misleading lung EIT imagesPhysiological Measurement, 2013
- GREIT: a unified approach to 2D linear EIT reconstruction of lung imagesPhysiological Measurement, 2009
- Bedside estimation of recruitable alveolar collapse and hyperdistension by electrical impedance tomographyIntensive Care Medicine, 2009
- Uses and abuses of EIDORS: an extensible software base for EITPhysiological Measurement, 2006
- A Matlab toolkit for three-dimensional electrical impedance tomography: a contribution to the Electrical Impedance and Diffuse Optical Reconstruction Software projectMeasurement Science and Technology, 2002