Trans-stenotic pressure drop estimation from PC-MRI and ultrasound imaging velocimetry using a modified Bernoulli equation
Open Access
- 31 July 2026
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 16 (1)
Abstract
Accurate non-invasive estimation of trans-stenotic pressure drops remains a challenge. In clinical practice, pressure drops are often estimated from velocity measurements using Bernoulli-based formulas, but these simplified relations do not explicitly account for how pressure losses change with the flow regime. Here, we introduce a modified Bernoulli (MB) formulation that incorporates regime-dependent pressure losses through a Reynolds-number-dependent loss coefficient. Steady in-vitro experiments were performed in an idealized stenosis model over physiologically relevant flow rates (0.65–3.9 L/min), combining direct pressure measurements with ultrasound imaging velocimetry (UIV) and phase-contrast magnetic resonance imaging (PC-MRI) to measure velocities. The MB model was calibrated from the measured pressure drops and then evaluated against the simplified Bernoulli (SB) and extended Bernoulli (EB) formulations. Over the tested flow regime, MB agreed best with the measurements (typically within about ±10% ). SB and EB showed larger biases, with errors of roughly 10–55% (SB) and -15 to 25%−(EB), and overestimated the pressure drop in the clinically relevant range. We additionally quantified the effect of PC-MRI in-plane pixel size on MRI-based pressure estimates. Coarse in-plane pixel size of 1.33 mm, corresponding to only approximately 3–4 pixels across the 3.3 mm throat radius of the 6.6 mm stenosis throat, led to systematic underestimation of flow rate and bulk velocity (about − 34 to − 44%) and, consequently, of the MB-predicted pressure drop (about − 52 to − 62%). In contrast, the peak throat velocity was substantially less sensitive to pixel size, resulting in smaller estimation errors (about − 13 to − 18.7%) when used as input for the MB. Overall, the results demonstrate that accounting for flow-regime-dependent loss mechanisms enhances pressure drop estimation, and that sufficient sampling of the stenotic throat is crucial for MRI-based flow rate and pressure drop estimation. In addition, peak-velocity-based MB pressure drop estimations are less sensitive to pixel size.Keywords
This publication has 53 references indexed in Scilit:
- A Statistical Shape Modeling Approach for the Derivation of a Data‐Driven Geometry‐Aware Lumped Arterial Stenosis ModelInternational Journal for Numerical Methods in Biomedical Engineering, 2026
- Investigation of the Pressure Drop in Arterial Models With Stenoses Using Numerical and Experimental In Vitro Approaches: Effect of ElasticityInternational Journal for Numerical Methods in Biomedical Engineering, 2025
- 4Dflow‐VP‐Net: A deep convolutional neural network for noninvasive estimation of relative pressures in stenotic flows from 4D flow MRIMagnetic Resonance in Medicine, 2023
- The effect of turbulence on transitional flow in theFDA's benchmark nozzle model using large‐eddy simulationInternational Journal for Numerical Methods in Biomedical Engineering, 2020
- Simplified Bernoulli's method significantly underestimates pulmonary transvalvular pressure dropJournal of Magnetic Resonance Imaging, 2015
- 4D Flow MRI‐based pressure loss estimation in stenotic flows: Evaluation using numerical simulationsMagnetic Resonance in Medicine, 2015
- Vascular Stenosis: An Introduction2015
- Echocardiography2012
- 4D flow MRIJournal of Magnetic Resonance Imaging, 2012
- Overestimation of aortic valve gradient measured by Doppler echocardiography in patients with aortic stenosisCatheterization and Cardiovascular Interventions, 2005