Baseline pressure errors (BPEs) extensively influence intracranial pressure scores: results of a prospective observational study
Open Access
- 1 January 2014
- journal article
- research article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 13 (1) , 7
Abstract
Monitoring of intracranial pressure (ICP) is a cornerstone in the surveillance of neurosurgical patients. The ICP is measured against a baseline pressure (i.e. zero - or reference pressure). We have previously reported that baseline pressure errors (BPEs), manifested as spontaneous shift or drifts in baseline pressure, cause erroneous readings of mean ICP in individual patients. The objective of this study was to monitor the frequency and severity of BPEs. To this end, we performed a prospective, observational study monitoring the ICP from two separate ICP sensors (Sensors 1 and 2) placed in close proximity in the brain. We characterized BPEs as differences in mean ICP despite near to identical ICP waveform in Sensors 1 and 2. The study enrolled patients with aneurysmal subarachnoid hemorrhage in need of continuous ICP monitoring as part of their intensive care management. The two sensors were placed close to each other in the brain parenchyma via the same burr hole. The monitoring was performed as long as needed from a clinical perspective and the ICP recordings were stored digitally for analysis. For every patient the mean ICP as well as the various ICP wave parameters of the two sensors were compared. Sixteen patients were monitored median 164 hours (ranges 70 – 364 hours). Major BPEs, as defined by marked differences in mean ICP despite similar ICP waveform, were seen in 9 of them (56%). The BPEs were of magnitudes that had the potential to alter patient management. Baseline Pressure Errors (BPEs) occur in a significant number of patients undergoing continuous ICP monitoring and they may alter patient management. The current practice of measuring ICP against a baseline pressure does not comply with the concept of State of the Art. Monitoring of the ICP waves ought to become the new State of the Art as they are not influenced by BPEs.Keywords
This publication has 23 references indexed in Scilit:
- Electrostatic discharges and their effect on the validity of registered values in intracranial pressure monitorsJournal of Neurosurgery, 2013
- Analysis of Intracranial PressureThe Neuroscientist, 2013
- A Trial of Intracranial-Pressure Monitoring in Traumatic Brain InjuryThe New England journal of medicine, 2012
- Simultaneous monitoring of static and dynamic intracranial pressure parameters from two separate sensors in patients with cerebral bleeds: comparison of findingsBioMedical Engineering OnLine, 2012
- A Randomized and Blinded Single-Center Trial Comparing the Effect of Intracranial Pressure and Intracranial Pressure Wave Amplitude-Guided Intensive Care Management on Early Clinical State and 12-Month Outcome in Patients With Aneurysmal Subarachnoid HemorrhageNeurosurgery, 2011
- Intracranial pressure: why we monitor it, how to monitor it, what to do with the number and what's the future?Current Opinion in Anaesthesiology, 2011
- The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic dischargesBioMedical Engineering OnLine, 2011
- Simultaneous measurements of intracranial pressure parameters in the epidural space and in brain parenchyma in patients with hydrocephalusJournal of Neurosurgery, 2010
- A dynamic nonlinear relationship between the static and pulsatile components of intracranial pressure in patients with subarachnoid hemorrhageJournal of Neurosurgery, 2010
- MULTICENTER CLINICAL ASSESSMENT OF THE RAUMEDIC NEUROVENT-P INTRACRANIAL PRESSURE SENSORNeurosurgery, 2008