Chronic impedance spectroscopy of an endovascular stent-electrode array
- 5 July 2016
- journal article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 13 (4) , 046020
Abstract
Objective. Recently, we reported a minimally invasive stent-electrode array capable of recording neural signals from within a blood vessel. We now investigate the use of electrochemical impedance spectroscopy (EIS) measurements to infer changes occurring to the electrode–tissue interface from devices implanted in a cohort of sheep for up to 190 days. Approach. In a cohort of 15 sheep, endovascular stent-electrode arrays were implanted in the superior sagittal sinus overlying the motor cortex for up to 190 days. EIS was performed routinely to quantify viable electrodes for up to 91 days. An equivalent circuit model (ECM) was developed from the in vivo measurements to characterize the electrode–tissue interface changes occurring to the electrodes chronically implanted within a blood vessel. Post-mortem histological assessment of stent and electrode incorporation into the wall of the cortical vessels was compared to the electrical impedance measurements. Main results. EIS could be used to infer electrode viability and was consistent with x-ray analysis performed in vivo , and post-mortem evaluation. Viable electrodes exhibited consistent 1 kHz impedances across the 91 day measurement period, with the peak resistance frequency for the acquired data also stable over time. There was a significant change in 100 Hz phase angles, increasing from −67.8° ± 8.8° at day 0 to −43.8° ± 0.8° at day 91, which was observed to stabilize after eight days. ECM’s modeled to the data suggested this change was due to an increase in the capacitance of the electrode–tissue interface. This was supported by histological assessment with >85% of the implanted stent struts covered with neointima and incorporated into the blood vessel within two weeks. Conclusion. This work demonstrated that EIS could be used to determine the viability of electrode implanted chronically within a blood vessel. Impedance measurements alone were not observed to be a useful predictor of alterations occurring at the electrode tissue interface. However, measurement of 100 Hz phase angles was in good agreement with the capacitive changes predicted by the ECM and consistent with suggestions that this represents protein absorption on the electrode surface. 100 Hz phase angles stabilized after 8 days, consistent with histologically assessed samples. Significance. These findings demonstrate the potential application of this technology for use as a chronic neural recording system and indicate the importance of conducting EIS as a measure to identify viable electrodes and changes occurring at the electrode–tissue interface.Keywords
This publication has 47 references indexed in Scilit:
- Comprehensive characterization of tungsten microwires in chronic neurocortical implants2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2012
- Reach and grasp by people with tetraplegia using a neurally controlled robotic armNature, 2012
- Comprehensive Analysis of Tissue Preservation and Recording Quality from Chronic Multielectrode ImplantsPLOS ONE, 2011
- Optimal spacing of surface electrode arrays for brain–machine interface applicationsJournal of Neural Engineering, 2010
- Implanted neural electrodes cause chronic, local inflammation that is correlated with local neurodegenerationJournal of Neural Engineering, 2009
- Signal quality of simultaneously recorded invasive and non-invasive EEGNeuroImage, 2009
- Cortical control of a prosthetic arm for self-feedingNature, 2008
- Electric Fields of the Brain2006
- Reliability of signals from a chronically implanted, silicon-based electrode array in non-human primate primary motor cortexIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2005
- Control of a two-dimensional movement signal by a noninvasive brain-computer interface in humansProceedings of the National Academy of Sciences of the United States of America, 2004